Table 5.

Intra-day and inter-day accuracy and precision for determination of totala hydroxybupropion (R,R and S,S) in human urine with analyte specific calibration rangesb on an AB Sciex 3200 mass spectrometer.

Intra-assay (n = 5)
Inter-assay (n = 5)
Analyte Nominal
concentration
(ng/mL)
Calculated
Concentration
(ng/mL)c
Accuracy (%) Precision (%) Calculated
Concentration
(ng/mL)b
Accuracy (%) Precision (%)
(R,R)-hydroxybupropion 100 91.3 ± 8.2 91 9 103 ± 6 103 6
500 517 ± 46 103 9 487 ± 43 97 9
2000 1983 ± 147 99 7 1870 ± 73 94 4
5000 5086 ± 350 102 7 4848 ± 129 97 3
10,000 9907 ± 536 99 5 9056 ± 337 91 4
15,000 15338 ± 381 102 2 14802 ± 774 99 5
(S,S)-hydroxybupropion 100 95.8 ± 13.7 96 14 93.9 ± 8.1 94 9
500 523 ± 36 105 7 472 ± 54 94 12
2000 2000 ± 96 100 5 1927 ± 74 96 4
5000 5346 ± 133 107 2 4976 ± 252 100 5
10,000 10004 ± 413 100 4 9343 ± 389 93 4
15,000 15082 ± 818 101 5 14592 ± 570 97 4
a

Total hydroxybupropion concentrations determined following sample preparation method 2 (Fig. 4).

b

Calibration ranges are provided in Table 1.

c

Mean ± standard deviation.