Results obtained for the evaluation dataset with the five commercial solutions [WFB (Mirada Medical), MIM (MIM Software), SPICE (Philips), ABAS (Elekta), and RS (RayStation)] compared to expert delineation for both femoral heads.

WFB MIM ABAS SPICE RS
Left femoral head R 0.93 ± 0.06 0.96 ± 0.13 0.96 ± 0.06 0.59 ± 0.06 0.98 ± 0.09
DSC mean 0.89 ± 0.05 0.89 ± 0.08 0.91 ± 0.04 0.70 ± 0.05 0.91 ± 0.03
DSC median 0.91 0.91 0.92 0.72 0.92
H95% (mm) 9.2 ± 6.4 9.9 ± 7.9 8.6 ± 6.9 29.7 ± 9.0 8.8 ± 7.2
Right femoral head R 0.93 ± 0.05 0.97 + 0.07 0.95 ± 0.05 0.60 ± 0.04 1.01 ± 0.07
DSC mean 0.91 ± 0.03 0.92 ± 0.02 0.92 ± 0.02 0.72 ± 0.03 0.92 ± 0.02
DSC median 0.92 0.93 0.92 0.72 0.92
H95% (mm) 8.1 ± 5.6 8.2 ± 5.3 8.5 ± 6.1 30.0 ± 6.5 6.4 ± 5.0

R is the volume ration, DSC is the Dice Similarity Coefficient, and H95% is the Hausdorff distance.