Skip to main content
. 2016 Jul-Sep;19(3):516–520. doi: 10.4103/0971-9784.185546

Table 3.

The reference values of thromboelastometer variables with differents rotational thromboelastometry tests in a multicentre study[7]

CT (s) CFT (s) MCF (mm) A10 (mm) CL 30 (%) ML (%)
INTEM 137-246 40-100 52-72 44-68 94-100 0-12
EXTEM 42-74 46-148 49-71 43-65 95-100 0-18
FIBTEM 43-69 - 9-25 9-24 - -

CT: Clot time, CFT: Clot formation time, MCF: Maximum clot firmness, A10: Amplitude 10 min after CT, CL30: Clot lysis index 30 min after CT, ML: Maximal clot lysis