Table 2. ESI-MS/MS parameters, concentration ranges (ng/mL), limits of detection (LOD), limits of quantification (LOQ) and linearity values (R 2 ) for AFs.

AFs MW Q1 (m/z) Q3 (m/z) CE (e/V) DP (V) range (ng/mL) R2 LOD (μg/ kg) LOQ (μg/kg) RSD (%)
AFB1 312.3 313.3 285.3a 30 178 0.0502–10.4 0.9987 0.008 0.011 2.9
    313.3 241.0 47 166
AFB2 314.3 315.3 287.1a 33 161 0.0350–7.0 0.9992 0.015 0.023 3.5
    315.3 259.0 38 159
AFG1 328.3 329.2 311.2a 30 143 0.0295–11.8 0.9985 0.022 0.029 4.6
    329.2 243.1 34 158
AFG2 330.3 331.2 217.0a 46 131 0.0295–11.8 0.9991 0.020 0.027 3.4
    331.2 245.3 38 114

aQuantitative ion.