Lesion and liver activities measured within the phantom calculated using the partition model and compared to the true activity measured at preparation

Activity (MBq)
Bremsstrahlung PET Tc-99m SPECT True
40-mm lesion 30.8 51.3 50.7 56.9
40-mm shell lesion 21.3 37.2 33.7 41.2
20-mm lesion 2.6 7.0 5.8 6.9
Liver 590.4 415.6 555.2 500.0