Skip to main content
. 2004 Aug;42(8):3739–3746. doi: 10.1128/JCM.42.8.3739-3746.2004

TABLE 1.

Reproducibility of EVs for the TaqMan HCV ASR calibratorsa

HCV RNA (IU/ml) No. detected/no. tested Mean EV SD % CV
4,000,000 11/11 22.97 0.272 1.19
400,000 11/11 26.38 0.199 0.75
40,000 11/11 29.72 0.252 0.85
4,000 11/11 32.62 0.204 0.63
400 11/11 34.60 0.167 0.48
40 11/11 37.21 0.599 1.61
4 8/11 40.8 0.976 2.47
a

Eleven replicates at each concentration were tested in three separate runs.