TABLE 1.
Reproducibility of EVs for the TaqMan HCV ASR calibratorsa
| HCV RNA (IU/ml) | No. detected/no. tested | Mean EV | SD | % CV |
|---|---|---|---|---|
| 4,000,000 | 11/11 | 22.97 | 0.272 | 1.19 |
| 400,000 | 11/11 | 26.38 | 0.199 | 0.75 |
| 40,000 | 11/11 | 29.72 | 0.252 | 0.85 |
| 4,000 | 11/11 | 32.62 | 0.204 | 0.63 |
| 400 | 11/11 | 34.60 | 0.167 | 0.48 |
| 40 | 11/11 | 37.21 | 0.599 | 1.61 |
| 4 | 8/11 | 40.8 | 0.976 | 2.47 |
Eleven replicates at each concentration were tested in three separate runs.