TABLE 2.

Reproducibility and accuracy of the TaqMan HCV ASR determined with a standard reference panel of HCV RNA

Labeled concn (IU/ml) Measured concna (IU/ml) SD CV (%) Bias (log10)
2,000,000 6,950,000 862,300 12 0.54
500,000 1,050,000 159,400 15 0.32
200,000 538,000 70,220 13 0.43
50,000 109,000 3610 3 0.34
5,000 12,200 1940 16 0.39
500 1,500 390 26 0.48
50 6 9 53 −0.92
a

Each concentration was tested in quadruplicate. HCV RNA was detected in all of the replicates.