Steps to optimize inhaled drug delivery during mechanical ventilation
| Step | Instruction(s) |
|---|---|
| 1 | Review inhaled-antibiotic order and assess need for pretreatment bronchodilator |
| 2 | Clear excess secretions from airway, if present |
| 3 | Load drug in nebulizer to manufacturer-recommended specifications |
| 4 | Position nebulizer on inspiratory line 10–15 cm from Y piece |
| 5 | Use continuous flow during nebulizer operation and remove heat-moisture exchanger from circuit |
| 6 | Set gas flow to nebulizer according to the manufacturer's specifications, and adjust ventilator limits to compensate for added flow |
| 7 | Run nebulizer until it begins to sputter |
| 8 | Remove nebulizer from circuit, rinse with sterile water, and run dry |
| 9 | Reconnect heat-moisture exchanger, and return ventilator to previous settings |
| 10 | Monitor patient for adverse events |