Skip to main content
. 2016 Aug 11;6:63. doi: 10.1186/s13550-016-0218-3

Table 3.

Precision and accuracy of plasma thymidine assay; acceptance criteria ±15 % for CV and RE (±20 % at LLOQ)

QC Theoretical thymidine concentration (ng/mL) Calculated thymidine mean concentration (ng/mL, n = 6) CV (%) RE (%)
LLOQ 1 0.99a 19 −0.9
Low 3 2.81 10 −6.4
Medium 40 47 13 −5.4
High 450 462 1.7 2.6

LLOQ Lower limit of quantification, CV coefficient of variation, RE relative error

a n = 5