Skip to main content
. Author manuscript; available in PMC: 2017 Sep 1.
Published in final edited form as: J Magn Reson Imaging. 2016 Mar 7;44(3):610–619. doi: 10.1002/jmri.25214

Table 3.

3.0 T NMR measured properties of the phantom components at 20 °C

Material components T1 relaxation time (ms) T2 relaxation time (ms)
Fibroglandular tissue mimic a 35% Corn syrup w/w in water 1230 ± 34.6 47.3 ± 0.9
Adipose tissue T1 and spectral shift mimic a Grapeseed oil 297 ± 6.3 32.5 ± 1.1
Diffusion mimics b
Benign lesion
10% PVP w/w in water 2300 ± 5.2 1953.1 ± 0.9
Malignant lesion 14% PVP 2010 ± 3.8 1692 ± 1.3
Malignant mass 18% PVP 1770 ± 3.4 1479.3 ± 0.9
25% PVP 1200 ± 3.4 1017.3 ± 0.8
40% PVP 714 ± 1.7 584.8 ± 0.2
a

T2 measurements made using spin-echo technique; details in the Methods section.

b

T2 measurements made using CPMG technique; details in the Methods section.