Skip to main content
. 2015 Nov 5;88(1056):20150389. doi: 10.1259/bjr.20150389

Table 2.

Image noise and subjective image quality

Protocol Reconstruction Image noise p-value Subjective image quality
Reader 1 Reader 2
Standard FBP 177 (155–193) p < 0.001 3 3
ADMIRE 3 99 (86–109) 4 4
ADMIRE 5 48 (42–52) 5 5
1/8 standard dose FBP 313 (297–320) p < 0.001 4 4
ADMIRE 3 182 (176–187) 4 4
ADMIRE 5 89 (88–91) 5 5
1/20 standard dose FBP 314 (305–321) p < 0.001 2 2
ADMIRE 3 186 (181–192) 3 3
ADMIRE 5 93 (88–97) 3 3
1/70 standard dose FBP 314 (305–321) p < 0.001 2 2
ADMIRE 3 196 (181–218) 3 3
ADMIRE 5 113 (100–133) 3 3

ADMIRE, advanced modelled iterative reconstruction; FBP, filtered back projection.

Image quality rated on a 5-point Likert scale.

Ranges are displayed in parentheses.