Table 1.

Baseline characteristics of the included studies

Study characteristics
Image quality
Routine dose group
Reduced dose group
Study Total number of patients IR technique Noise CNR SNR Subjective image quality BMI (routine dose group, kg m−2) Number of patients Tube current Tube voltage (kV) Mean effective dose (normal, mSv) BMI (reduced dose group, kg m−2) Number of patients Tube current Tube voltage (kV) Mean effective dose (reduced, mSv)
Renker et al14 24 IRIS Somatom® Definition Flash + NR NR + 30.7 12 NR 120 9.6 30.2 12 NR 80/100 3.7
Wang et al15 78 SAFIRE Somatom Definition Flash ± ± ± ± 31.7 39 354–430 mAs 120 8.8 32.3 39 286–370 mAs 100 4.4
Tomizawa et al16 100 AIDR Aquilion One™ + ± ± ± 24.0 50 483 mA 120 5.5 23.9 50 289 mA 120 4.2
Hou et al13 110 iDose Brilliance iCT ± ± ± ± 25.1 21 210 mAs 120 3.2 25.5 16 65 mAs 120 0.9
Gebhard et al17 70 ASIR Discovery™ HD 750 + ± ± + 33.8 35 646 mA 116 2.5 32.9 35 633 mA 112 2.3
Yin et al18 60 SAFIRE Somatom Definition Flash ± ± ± ± NR 60 320–400 mAs 80–120 1.5 NR 60 160–200 mAs 80–120 0.7
Carrascosa et al19 200 iDose Brilliance iCT ± NR NR ± 27.2 100 203 mAs 119 4.6 26.3 100 196 mAs 109 2.8
Takx et al20 20 SAFIRE Somatom Definition Flash 33.1 20 331 mAs 100–120 5.5 33.1 20 66 mAs 100–120 1.1
Shen et al21 338 ASIR Discovery HD 750 ± NR + 25.5 109 600 mA 120 2.9 25.2 116 300 mA 120 1.8
Fuchs et al22 42 ASIR/MBIR Discovery HD 750 + NR + ± 25.2 42 450–700 mA 100–120 1.2 25.2 42 150–210 mA 80–100 0.2

−, decreased with IR; +, improved with IR; ±, no difference; AIDR, Adaptive Iterative Dose Reduction; BMI, body mass index; CNR, contrast-to-noise ratio; IR, iterative reconstruction; IRIS, Iterative Reconstruction in Image Space; MBIR, Model-Based Iterative Reconstruction; NR, not reported; SAFIRE, Sinogram-Affirmed Iterative Reconstruction; SNR, signal-to-noise ratio.

The Somatom Definition Flash was obtained from Siemens Medical Systems, Forchheim, Germany; Aquilion One™ was obtained from Toshiba Medical Systems Co. Ltd, Ottowara, Japan; Brilliance iCT was obtained from Philips Healthcare, Best, Netherlands; Discovery™ HD 750 was obtained from GE Healthcare, Milwaukee, WI.