Table 9.

Average mean glandular dose (MGD) to the 53 mm-thick standard breast model for different types of digital radiography (DR), computed radiography (CR) and film-screen systems

Manufacturer and model Number of systems Mean MGD to standard breast (mGy) (±2SEM)
Fujifilm Amulet 7 1.06 ± 0.08
GE DS 26 1.19 ± 0.03
GE Essential 40 1.24 ± 0.04
Hologic Selenia® Mo 3 2.06 ± 0.25
Hologic Selenia® W 45 1.43 ± 0.04
Hologic Dimensions® 27 1.46 ± 0.02
IMS Giotto 2 1.56 ± 0.24
Philips MicroDose 16 0.68 ± 0.03
Siemens Inspiration 47 1.10 ± 0.06
All DR systems 214 1.24 ± 0.04
Fujifilm Profect CR 9 2.15 ± 0.26
Film-screen systems 160 1.67 ± 0.06
All systems 359 1.43 ± 0.04

SEM, standard error in the mean.

Fischer (Hologic, Inc., Marlborough, MA); Fujifilm (FUJIFILM UK Ltd, Bedford, UK); GE (GE Medical Systems, Buc, France); Hologic (Hologic, Inc.); IMS (Internazionale Medico Scientifica, Bologna Italy); Planmed (Planmed Oy, Helsinki, Finland); Philips (Philips Healthcare, Guildford, UK); Siemens (Siemens Healthcare GmbH, Erlangen, Germany).