Skip to main content
. 2016 Mar 28;89(1061):20150705. doi: 10.1259/bjr.20150705

Table 1.

Summary of scanning parameters for different types of CT scanners

Single source
 64- and 80-detector row scanners (including “128- and 160-slice” systems)
     HR < 65 Prospective gating with end-diastolic acquisition
     HR > 65 Prospective gating with end-systolic acquisition
     Exception Where phase selection is not adjustable (e.g. on a scanner with prospective helical acquisition with diastolic phase acquisition only for slow HRs)
There may be a role for retrospective gating (e.g. when the HR is >100 beats per minute)
When retrospective acquisition is used, dose modulation outside the 30–80% cardiac cycle should be applied
For scanners that have a retrospective mode with adaptive dose modulation, this may be used as an alternative for fast HRs
 128-, 256- or 320-detector row scanners (including “256- and 640-slice” systems)
  128–256 detector rows
     HR < 75 Prospective gating with end-diastolic acquisition
     HR > 75 Prospective gating with end-systolic acquisition
  320 detector rows
     HR independent Non-gated helical acquisition with the middle 8-cm coverage (160 × 0.5 mm) can be used to image the thoracic aorta in 1–2 heartbeats with motion-free imaging of the aorta
     Exception If dedicated coronary assessment is required (e.g. in the context of known AAS or a high pre-test probability), then use following
     HR < 65 Prospectively triggered ECG synchronization with 70–80% single pulse per volume
     HR > 65 Prospectively triggered ECG synchronization with 30–80% single pulse per volume
Dual source
          HR-dependent HR-dependent prospectively ECG-synchronization protocols can be applied similar to the systems above
          HR < 65 Prospective gating with end-diastolic acquisition
          HR > 65 Prospective gating with end-systolic acquisition
  In a system that allows for high-pitch acquisition in conjunction with wide detector arrays, traditional ECG synchronization may not be required
e.g. a pitch of >3 and gantry rotation time 0.28 s permit coverage of 9.6–11.6 cm s−1

AAS, acute aortic syndrome; ECG, electrocardiogram; HR, heart rate.