. 2016 Feb 1;89(1059):20150715. doi: 10.1259/bjr.20150715

Table 2.

Vertical paddle motion for small paddles (18 × 24 cm) during the first, second, third and fourth section of 10-s time periods after clamping commencement

Time period (s)
0–10
10–20
20–30
30–40
0–40
Paddle type Average paddle motion ( ± SD, n) (mm/10 s)
GE Senographe Essential (flexible) 0.21±0.06, 120 0.08±0.03, 120 0.04±0.01, 120 0.03±0.01, 120 0.36±0.09, 480
Hologic Lorad Selenia (flexible) 0.26±0.07, 20 0.05±0.01, 20 0.03±0.01, 20 0.03±0.01, 20 0.37±0.08, 80
GE Senographe Essential (fixed) 0.26 ± 0.07, 120 0.06 ± 0.02, 120 0.05 ± 0.01, 120 0.02 ± 0.01, 120 0.39 ± 0.09, 480
Siemens Mammomat Inspiration (fixed) 0.28 ± 0.08, 20 0.13 ± 0.04, 20 0.08 ± 0.02, 20 0.05 ± 0.02, 20 0.54 ± 0.14, 80
Siemens Mammomat Inspiration (flexible) 0.35±0.11, 20 0.13±0.03, 20 0.10±0.02, 20 0.05±0.01, 20 0.63±0.16, 80
Hologic Selenia Dimensions (flexible) 0.39±0.12,80 0.18±0.05,80 0.12±0.04,80 0.10±0.03,80 0.79±0.22,320
Hologic Selenia Dimensions (fixed) 0.51 ± 0.15, 80 0.18 ± 0.05, 80 0.11 ± 0.03, 80 0.07 ± 0.02, 80 0.87 ± 0.22, 320

is the mean; SD, standard deviation; n, number of observations.

Flexible paddles are in italicized text.

Hologic Lorad Selenia, Hologic Selenia is obtained from Hologic, Bedford, MA; GE Senographe Essential is obtained from General Electric Medical Systems, Milwaukee, WI; and Siemens Mammomat Inspiration is obtained from Siemens, Erlangen, Germany.