. 2016 Feb 1;89(1059):20150715. doi: 10.1259/bjr.20150715

Table 3.

Vertical paddle motion for large paddles (24 × 29 cm and 24 × 30 cm) during the first, second, third and fourth 10 s time periods after clamping commencement

Time period (s)
0–10
10–20
20–30
30–40
0–40
Paddle type Average paddle motion ( ± SD, n) (mm/10 s)
Hologic Lorad Selenia (flexible) 0.17±0.05, 20 0.06±0.02, 20 0.03±0.01, 20 0.01±0.01, 20 0.27±0.07, 80
GE Senographe Essential (flexible) 0.30±0.09, 80 0.06±0.02, 80 0.05±0.02, 80 0.04±0.01, 80 0.45±0.10, 320
GE Senographe Essential (fixed) 0.31 ± 0.09, 100 0.08 ± 0.02, 100 0.04 ± 0.01, 100 0.03 ± 0.01, 100 0.46 ± 0.10, 400
Siemens Mammomat Inspiration (flexible) 0.33±0.10, 20 0.12±0.04, 20 0.09±0.03, 20 0.04±0.01, 20 0.58±0.15, 80
Hologic Selenia Dimensions (flexible) 0.35±0.11, 80 0.15±0.04, 80 0.10±0.03, 80 0.05±0.02, 80 0.65±0.17, 320
Hologic Selenia Dimensions (fixed) 0.42 ± 0.13, 80 0.13 ± 0.04, 80 0.07 ± 0.02, 80 0.06 ± 0.02, 80 0.68 ± 0.16, 320

is the mean; SD, standard deviation; n, number of observations.

Flexible paddles are in italicized text.

Hologic Lorad Selenia, Hologic Selenia is obtained from Hologic, Bedford, MA; GE Senographe Essential is obtained from General Electric Medical Systems, Milwaukee, WI; and Siemens Mammomat Inspiration is obtained from Siemens, Erlangen, Germany.