Skip to main content
. 2016 Aug;6(4):358–368. doi: 10.1212/CPJ.0000000000000265

Figure 3. Longitudinal midline view.

Figure 3

The transducer is centered over a lumbar interspinous space in a longitudinal plane, and a mark is made perpendicular to the center of the transducer. The spinous processes (SP) and interspinous spaces (*) are visualized in a longitudinal plane, and the ligamentum flavum (LF) and posterior longitudinal ligament (PLL) are visualized deep to the spinous processes.