Limits of the axial coordinate estimation accuracy for a conventional microscopy
setup and a two-plane MUM setup. The limit of accuracy
δz0 for the
estimation of the z0 coordinate of an out-of-focus
and stationary molecule from a conventional image (*) or a pair of
two-plane MUM images (○) is plotted as a function of the coordinate
z0. Focal planes 1 and 2 of the MUM setup are
denoted by vertical dashed lines, and are located at
z0 = 0 nm and
z0 = 450 nm, respectively. The value of
z0 is specified with respect to focal plane 1,
which coincides with the focal plane of the conventional setup. In both setups,
an image consists of an 11×11-pixel region, acquired using a CCD
detector with a pixel size of 13 μm × 13
μm and readout noise with mean
η0 = 0 electrons and standard
deviation σ0 = 6 electrons at each
pixel. The image of the molecule is described by the Born and Wolf image
function. For the conventional setup, the molecule is laterally positioned such
that the center of its image is located at 5.05 pixels in the x
direction and 5.25 pixels in the y direction. Photons are
detected from the molecule at a rate of Λ0 = 10000
photons/s over an acquisition time of t −
t0 = 0.1 s, such that
Λ0 · (t −
t0) = 1000 photons are on average
detected from the molecule over the detector plane (and 952.3 photons are on
average detected from the molecule over a given image when the molecule is in
focus). A background detection rate of is assumed, along with a uniform background
spatial distribution of fb(x,
y) = 1/20449
μm−2, such that there is a
background level of β0 = 20
photoelectrons per pixel over the 11×11-pixel image. Photons of
wavelength λ = 520 nm are collected by an
objective lens with magnification M = 100 and numerical
aperture na = 1.4. The refractive index of
the immersion medium is n = 1.518. For the MUM setup,
the collected molecule and background photons are split equally between the
images corresponding to the two focal planes. Therefore, the photon detection
rate is Λ0 = 5000 photons/s per image, and the
background detection rate is per image. For the image corresponding to focal
plane 1, all other details are as given for the conventional setup. For the
image corresponding to focal plane 2, the magnification is M
= 98.18, calculated using a formula from [76] for a plane spacing of
Δzf = 450 nm and a tube
length of 160 mm. The lateral position of the image of the molecule and the
background spatial distribution are accordingly adjusted based on this slightly
smaller magnification.