. 2016 Aug 16;43(9):5117–5130. doi: 10.1118/1.4960632

TABLE V.

The mean percentages of pixels passing for gamma analysis (±2%/2 mm criteria) comparing film-measured dose to calculated dose for two repeated film measurements (with metal implants in place). Comparisons were performed for the four possible combinations of the two error-reduction strategies (CT AR methods and MK).

Mean % pixels passing
Phantom AR method Baselinea AR Baseline + MK AR + MK
Spine phantom o-mar 81.0 77.9 82.9 80.2
gsi 140 keV 74.3 62.4 80.7 60.3
mars 140 keV 74.3 43.5 80.7 46.4
Dental phantom o-mar 69.8 72.0 70.5 71.8
gsi 140 keV 68.1 67.3 69.0 66.7
mars 140 keV 68.1 71.6 69.0 70.4
a

Baseline imaging was Philips 120 kVp for o-mar and GE 120 kVp for both gsi 140 keV and mars 140 keV imaging.