Skip to main content
. 2016 Aug 22;16(1):48. doi: 10.1186/s12880-016-0155-7

Table 2.

Area under the receiver operating characteristic curve values obtained from MRMCROC method for three radiologists in predicting lung nodule calcification on conventional gray scale radiography (GSR) and dual-energy radiography with color-representation (DCR)

Observer Gray scale radiography Dual-energy radiography with color-representation
1 0.871 0.949
2 0.739 0.957
3 0.874 0.926
Average 0.828 0.944

The difference in area under the receiver operating characteristic curve values between GSR and DCR imaging was statistically significant (p = 0.044)

MRMCROC multi-reader, multi-case receiver operating characteristic