Table 1.
Signals Measured from Two 35 mm3 VOIs Placed Side By Side Inside a Homogeneous Phantom Close to the Coil
Shared excitation slice | Area of signal peak (a.u) | SNR | Signal loss |
---|---|---|---|
Left voxel reference | 6563 ± 49 | 27.2 ± 1.1 | – |
Right voxel reference | 7693 ± 48 | 22.8 ± 1.1 | – |
Left voxel MVS shared excitation | 3527 ± 12 | 12.5 ± 0.4 | 46 ± 0.3% |
Right voxel MVS shared excitation | 4991 ± 55 | 10.6 ± 0.4 | 35 ± 1% |
Shared refocusing slices | |||
Left voxel reference | 3994 ± 22 | 17.7 ± 0.6 | – |
Right voxel reference | 6178 ± 55 | 17.7 ± 0.8 | – |
Left voxel MVS (1 + 7) | 3607 ± 19 | 16.1 ± 0.6 | 10 ± 1% |
Right voxel MVS (1 + 7) | 4860 ± 50 | 14.7 ± 0.6 | 21 ± 1% |
Left voxel MVS (4 + 4) | 3737 ± 29 | 16.7 ± 0.8 | 6 ± 1% |
Right voxel MVS (4 + 4) | 5563 ± 56 | 16.4 ± 0.7 | 10 ± 1% |
Left voxel MVS (4 + 4) + 12.5% | 3880 ± 18 | 16.7 ± 0.7 | 3 ± 1% |
Right voxel MVS (4 + 4) + 12.5% | 5681 ± 44 | 16.5 ± 0.4 | 8 ± 1% |
Both VOIs were acquired separately with T R = 8 s as reference signals. For multivoxel spectroscopy (MVS), the delay between the localizations was 1 s, followed by a delay of 7 s with a shared excitation slice only and vice versa for each voxel position. The influence of sharing both refocusing pulses was tested with asymmetric (1 + 7 s) and symmetric (4 + 4 s) timing. To reduce saturation effects caused by refocusing pulses, the measurement was repeated with the refocusing pulse voltage increased by 12.5% (from 160 to 180 V).