Skip to main content
. 2016 Aug 25;11(8):e0161721. doi: 10.1371/journal.pone.0161721

Table 3. Acceptable and desired target product profile characteristics focused on operational characteristics, as defined by an expert consensus process 2015–16.

Characteristic Acceptable ("must have") Desired ("would like") Reference
Multiplexing ≥1 analyte ≥1 analyte plus pathogen specific testing (priority: malaria) Expert consensus
Ease of test performance ≤2 timed steps during assay performance No timed step during the assay performance Expert consensus
Sample type Capillary blood or urine Capillary blood or any less invasive sample than blood, like: Saliva Expert consensus
Volume - Capillary blood:
- patients > 6 years[41] - 100μL (~ 4 drops);
- pediatric patients 2–6 years[41]– 50–100μL (~2 drops);
- Saliva: < 0.5mL
- Urine: ~1mL of urine
- Capillary blood: 25μL (~1 drop) for all age groups
- Saliva: <0.5mL for saliva
Expert consensus
Sample collection Transfer and quantification device included in the test Industry standard
Additional sample preparation 1 sample-processing steps None required Expert consensus
Kit configuration No additional reagents outside of the kit required (including gloves) Industry standard
Process control Internal control to provide test validity and acceptance Expert consensus
Batch/Quality control Positive and negative controls required to monitor the quality of kit Expert consensus
Reagent preparation - Minimal of one additional step to prepare prior to use
- No (precise) measuring required.
No additional reagents required, everything is provided ready-to-use [32]
Time to result (per sample) <2 hours <10min Expert consensus, Ivanova et. al. in preparation
Hands on time Total hands-on-time should be <5 min Total hands on time should be <1 min Expert consensus
Sample throughput Ability to test individual samples or multiple samples if needed (no need for batching) Expert consensus
Result stability ≥15 min ≥1 hour Industry standard
Assay type Qualitative Quantitative Expert consensus
Biosafety No need for a biosafety cabinet; basic safety procedures need to be followed [42]
Waste disposal Biohazard waste - Testing device
- Compostable plastics for minimal environmental impact
[42]
Storage conditions and Self-life - 12 month at fluctuating temperature (0–40°C)
- ≤90% relative humidity
- No controlled temperature required
- 24 month at fluctuating temperature (0–50°C)
- ≤90% relative humidity
- No controlled temperature required
Expert consensus
Operation conditions - Between 5°C—40°C
- ≤90% humidity
- Between 5°C—45°C
- ≤90% humidity
Expert consensus
Shipping conditions Shipping without cold chain; should tolerate stress during transport (≤72h at +50°C) Expert consensus
Training requirementsa <2 days including proficiency panel <0.5 day including proficiency panel [32]
Equipment (instrument external to test) Small, robust, dust-resistant, portable or hand-held integrated instrument that must operate on battery Reusable instrumentation not required Expert consensus
Power supply Battery or solar powered None required [32]
Water supply No external water required Industry standard
External maintenance - Preventative maintenance at 2 year or >4000 samples; simple with only minimal expertise
- Maintenance alert should be included.
None required Industry standard
Calibration Remote calibration or auto-calibration No calibration required Industry standard
Data output Qualitative Non-ambiguous results displayed (e.g. bacterial/non-bacterial–yes or no)  Expert consensus
Data interpretationb Minimal interpretation required No interpretation required Expert consensus
Connectivity Not required Wireless connectivity Isaak et. al. in preparation
Reproducibility >95% standard deviation between repeats [43]

a Training to use the test, not including clinical/treatment implications and consequences.

b Interpretation of the test result, not the clinical/treatment conseque