Table 1.
Frequency | QA test | Tolerance |
---|---|---|
Daily | Imager positioning/repositioning | ±1 mm |
Imaging and treatment beam coincidence (single gantry angle) | ±1 mm | |
Monthly | Imaging and treatment beam coincidence (4 cardinal gantry angles) | ±1 mm |
Image scaling | ±2 mm | |
Mechanical alignment – full range of travel | ±2 mm | |
Geometric accuracy for CBCT | ±2 mm | |
Automatic image registration and offset calculation accuracy | ±2 mm |