Skip to main content
. 2016 Aug 26;17(1):120. doi: 10.1186/s12882-016-0315-6

Table 1.

Characteristics of HFR and HFR-S filters

Filter section HFR HFR-S
Membrane Type (surface) convective PES HF (0.7 m2) PES HHF (0.7 m2)
Sterilization mode: gamma rays diffusive PES LF (1.7 m2) PES LF (1.7 m2)
Kuf (mL/h/mmHg) convective 25 39
diffusive 13 13
Albumina sieving coefficient convective 0.002 0.2
diffusive <0.0001 <0.0001
Adsorbent cartridge (volume) Sterilization mode: dry steam SDVB (40 mL) SDVB (80 mL)

PES Polyethersulfone (LF Low flux, HF high flux, HHF Hyper high flux)

SDVB Styrene Divinylbenzene

aBovin Serum Albumin