Micro-CT scan acquisition protocol

Micro-CT scanner model Phoenix v|tome|x s 180/240 (dual tubes)
X-ray tube Phoenix 180kV/15W high-power nanofocus tube
Detector Temperature stabilized digital GE DXR detector array
X-ray source target Diamond
Isotropic voxel size (spatial resolution) Approx. 9.5μm
Detector panel resolution (width x height) in pixels (in inches) 2024 × 2024 (16″×16″)
Field of view (width x height) in mm 11.4 ×19.2
Rotation 360° in 33 min 22 sec
Projections 1000
Number of exposures (projections) 3
Exposure time 500 msec/projection
Voltage (kV)/current (mA) 90/110 and 100/91
Staining (soaking) agent Lugol’s Iodine, 1.25% w/v I2 and 2.5% w/v KI in deionized water