. 2016 Aug 24;54(9):2337–2342. doi: 10.1128/JCM.00897-16

TABLE 2.

Clinical performance of the Xpert HPV compared to a standard, clinically validated comparator test (GP5+/6+ EIA or hc2)

Age group Measure Outcome (No.) Absolute accuracy (% [95% CI])
Relative accuracy: Xpert HPV vs SCT (95% CI) PMcN Pn.inf
Xpert HPV SCT
VALGENT 2
    All ages Sensitivity CIN2+ (101) 94.1 (87.5–97.8) 94.1 (87.5–97.8) 1.000 (0.960, 1.042) 1.00 0.0017
Sensitivity CIN3+ (55) 98.2 (90.3–100) 98.2 (90.3–100) 1.000 1.00 0.0072
Specificity ≤CIN1 (842) 82.7 (79.9–85.1) 83.2 (80.6–85.7) 0.993 (0.972–1.014) 0.50 <0.0001
    ≥30 yr Sensitivity CIN2+ (41) 90.2 (76.9–97.3) 92.7 (80.1–98.5) 0.974 (0.889–1.066) 1.00 0.0951
Sensitivity CIN3+ (24) 95.8 (78.9–99.9) 95.8 (78.9–99.9) 1.000 1.00 0.055
Specificity ≤CIN1 (652) 88.3 (85.6–90.7) 89.4 (86.8–91.7) 0.988 (0.968–1.008) 0.24 <0.0001
Aggregated set
    All ages Sensitivity CIN2+ (180) 96.1 (92.2–984) 96.1 (92.2–98.4) 1.000 (0.978–1.023) 1.00 <0.0001
Sensitivity CIN3+ (102) 99.0 (94.7–100) 99.0 (94.7–100) 1.000 1.00 0.0004
Specificity ≤CIN1 (4,171) 82.5 (81.3–83.6) 82.1 (80.9–83.3) 1.001 (0.992–1.011) 0.76 <0.0001
    ≥30 yr Sensitivity CIN2+ (68) 92.6 (837–97.6) 94.1 (85.6–98.4) 0.984 (0.931–1.040) 1.00 0.019
Sensitivity CIN3+ (38) 97.4 (86.2–99.9) 97.4 (86.2–99.9) 1.000 1.00 0.021
Specificity ≤CIN1 (3,206) 88.1 (86.9–89.2) 87.5 (86.3–88.6) 1.006 (0.997–1.016) 0.18 <0.0001

a Absolute sensitivities of the tests for CIN2+ or CIN3+ and specificity for ≤CIN1 are presented, as is the relative accuracy of the Xpert HPV compared to the comparator tests. The statistical tests in the last two columns verify differences (P for the McNemar test [PMcN] or noninferiority [Pn.inf]). The upper six rows show accuracy measures according to the VALGENT-2 set where the comparator test was GP5+/6+ EIA. The aggregated data set incorporates VALGENT-2 and data from Cuzick et al. (7) where the comparator test was hc2.