Skip to main content
. 2016 Aug 17;2016:1234581. doi: 10.1155/2016/1234581

Table 6.

Performance data of DrugWipe 5A in comparison with HS-SPME-GC-MS results.

Cannabis Amphetamines Cocaine Opiates
TP 14 26 16 2
FP 4 7 3 7
TN 30 43 56 72
FN 35 7 8 2

Number of tests 83 83 83 83

Sensitivity (%) 29 79 67 50
Specificity (%) 88 86 95 91
Accuracy (%) 53 83 87 89