Skip to main content
. 2016 Sep 2;6:32615. doi: 10.1038/srep32615

Table 4. Quantitative estimates of the maximum electrode movement allowed in % positioning the electrodes in the x and y axes to achieve a pre-specified reproducibility given by the intra-class correlation (ICC).

ICC x-axis, %
y-axis, % All
HC, HP LP, LC
≥0.9 13.6 8.7 37.5
≥0.8 18.5 12.9 44.6
≥0.7 22.2 16.4 46.6
≥0.6 24.8 19.6 47.5
≥0.5 26.9 22.7 48.3
≥0.4 30.0 26.1 48.9
≥0.3 33.2 29.2 49.5
≥0.2 36.1 32.7 50.0

The maximum permissible electrode misplacement decreases with increasing the reproducibility. Electrodes nomenclature: HC, high current; HP, high potential; LP, low potential; LC, low current.