Table 3. Agreement of total retinal thickness between OCT devices.

Mean difference (bias; μm) and 95% limits of agreement (μm) for mean retinal thickness at session 1 produced by the Iowa Reference Algorithms, using AEL-dependent scaling, for each pairing of OCT instruments.

Fovea Inner ring Outer ring
1040nm-Topcon Mean difference -14.18 -7.05 0.80
Limits of agreement -20.05 to -8.30 -14.65 to 0.55 -6.95 to 8.55
Friedman test P 0.000* 0.000* 0.317
1040nm-Zeiss Mean difference -13.89 -12.46 -6.62
Limits of agreement -21.62 to -6.16 -18.91 to -6.00 -13.53 to 0.30
Friedman test P 0.000* 0.000* 0.000*
Topcon-Zeiss Mean difference -0.21 -5.41 -7.42
Limits of agreement -4.18 to 3.75 -8.67 to -2.15 -10.66 to -4.17
Friedman test P 0.841 0.000* 0.000*

*significant at 0.05 level.