Abstract
AIMS: To compare the International Normalised Ratio (INR) obtained directly with the two types of WHO plain International Reference Preparation for thromboplastin in patients treated with coumarin. METHODS: Prothrombin times were performed in parallel at four centres using WHO human plain IRP (BCT/253) and rabbit plain IRP (RBT/79). Sixty patients and 20 normal controls were tested at each centre. Differences in INR among the centres were assessed by one factor, analysis of variance. The bias for each centre was assessed by the t test. RESULTS: At all four centres higher INRs were consistently found with the rabbit plain reagent. Two of the centres showed significantly greater bias. CONCLUSIONS: There was a small but significant difference in INR results obtained directly with these two reference reagents at all four centres (mean 7.35%). This in part may result from the different responsiveness of the two IRP to the coumarin defect or to imprecision of the original ISI calibrations of the two plain WHO IRP. The findings support the adoption of a single master IRP, in accord with WHO recommendations, which would resolve the present anomalous situation.
Full text
PDF


Selected References
These references are in PubMed. This may not be the complete list of references from this article.
- Hermans J., van den Besselaar A. M., Loeliger E. A., van der Velde E. A. A collaborative calibration study of reference materials for thromboplastins. Thromb Haemost. 1983 Oct 31;50(3):712–717. [PubMed] [Google Scholar]
- Ingram G. I. The stability of the WHO reference thromboplastin NIBS & C 67/40. Thromb Haemost. 1979 Dec 21;42(4):1135–1140. [PubMed] [Google Scholar]
- Kirkwood T. B. Calibration of reference thromboplastins and standardisation of the prothrombin time ratio. Thromb Haemost. 1983 Jun 28;49(3):238–244. [PubMed] [Google Scholar]
- Poller L. Laboratory control of oral anticoagulants. Br Med J (Clin Res Ed) 1987 May 9;294(6581):1184–1184. doi: 10.1136/bmj.294.6581.1184. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Poller L., Taberner D. A. Dosage and control of oral anticoagulants: an international collaborative survey. Br J Haematol. 1982 Jul;51(3):479–485. doi: 10.1111/j.1365-2141.1982.tb02805.x. [DOI] [PubMed] [Google Scholar]
- Taberner D. A., Poller L., Thomson J. M., Darby K. V. Effect of international sensitivity index (ISI) of thromboplastins on precision of international normalised ratios (INR) J Clin Pathol. 1989 Jan;42(1):92–96. doi: 10.1136/jcp.42.1.92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Thomson J. M., Darby K. V., Poller L. Calibration of BCT/441, the ICSH reference preparation for thromboplastin. Thromb Haemost. 1986 Jun 30;55(3):379–382. [PubMed] [Google Scholar]
- Thomson J. M., Tomenson J. A., Poller L. The calibration of the second primary international reference preparation for thromboplastin (thromboplastin, human, plain, coded BCT/253). Thromb Haemost. 1984 Dec 29;52(3):336–342. [PubMed] [Google Scholar]