Gem |
TE = 6–15 ms |
Gradient echo sequence for PRF-shift thermometry. |
|
TR = 30 ms |
All monitoring was conducted in a single |
|
Angle = 25 |
slice in the MRI axial plane, parallel to the direction |
|
Mat = 96×96 |
of acoustic propagation. 1–2 dummy |
|
FOV = 60×60 mm |
scans were used to suppress steady-state artifacts. |
Fsems |
TE = 37 ms |
|
|
TR = 3000 ms |
Fast spin echo sequence for T
2-weighted anatomical |
|
ETL = 8 |
imaging. Enables tumor localization and visualization |
|
ESP = 9 ms |
of the surrounding environment. |
|
Angle = 20 |
|
Mgems |
TE = 2 ms |
|
|
ΔTE = 3 ms |
|
|
TR = 30 ms |
Multi-echo gradient echo scan for water/fat separation |
|
Angle = 25 |
in post-treatment analysis. |
|
Mat = 96×96 |
|
|
FOV = 60×60 mm |
|
Gems_meg |
TE = 7.6 ms |
Modified gradient echo sequence for ARFI. G_dur |
|
TR = 71 ms |
represents the duration a single lobe of the biopolar |
|
G_amp = 10 G/cm |
MEG. The direction of motion encoding was |
|
G_dur = 4 ms |
controlled within the scanner interface based on |
|
FUS = 1.1 MHz |
the slice orientation. |