Table 1.

MR imaging sequences and parameters

Sequence Parameters Purpose
Gem TE = 6–15 ms Gradient echo sequence for PRF-shift thermometry.
TR = 30 ms All monitoring was conducted in a single
Angle = 25 slice in the MRI axial plane, parallel to the direction
Mat = 96×96 of acoustic propagation. 1–2 dummy
FOV = 60×60 mm scans were used to suppress steady-state artifacts.
Fsems TE = 37 ms
TR = 3000 ms Fast spin echo sequence for T 2-weighted anatomical
ETL = 8 imaging. Enables tumor localization and visualization
ESP = 9 ms of the surrounding environment.
Angle = 20
Mgems TE = 2 ms
ΔTE = 3 ms
TR = 30 ms Multi-echo gradient echo scan for water/fat separation
Angle = 25 in post-treatment analysis.
Mat = 96×96
FOV = 60×60 mm
Gems_meg TE = 7.6 ms Modified gradient echo sequence for ARFI. G_dur
TR = 71 ms represents the duration a single lobe of the biopolar
G_amp = 10 G/cm MEG. The direction of motion encoding was
G_dur = 4 ms controlled within the scanner interface based on
FUS = 1.1 MHz the slice orientation.