Table 1.

Characteristics of tested devices, baseline characteristics of patients, and differences in usability

Breezhaler Genuair Handihaler p
Device characteristics
Manoeuvres (n)a 7 3 4
 Time spent by nurse with patients during trainingb (min) 1st attempt 2.8 ± 0.1 1.3 ± 0.1 1.4 ± 0.1 p < 0.0001
2nd attempt 4.2 ± 0.2 1.8 ± 0.1 2 ± 0.2 p < 0.0001
3rd attempt 5.0 ± 0.3 2.1 ± 0.2 2.7 ± 0.2 p < 0.0001
Patient characteristics
 Patients (N) 333 237 223
 Age (years) 55.2 ± 18.3 54.8 ± 17.5 57.8 ± 17.9 NS
 Sex (% male) 46.5 % 49.4 % 43.9 % NS
 Disease (% COPD) 47.4 % 47.7 % 51.6 % NS
 Previous experience with DPI 63.7 % 66.7 % 64.6 % NS
 Previous instruction to use of DPI 60.7 % 64.1 % 62.8 % NS
Usability
 N. attempts before achieving the first proper inhalation 2.6 ± 1.1 1.6 ± 0.8 2.5 ± 1.1 p < 0.0001
 Successful rate at the first attempt (%) 18.0 % 55.7 % 19.3 % p < 0.0001
 Total time required for the first proper inhalationb (min) 10.3 ± 5.0 2.5 ± 1.6 4.9 ± 2.7 p < 0.0001

DPI dry powder inhaler

aNumber of operations to prepare the inhalation

bIncluding patient time and nurse demonstration