Table 2.

Predictors of proper inhalation achieving: results of univariate and multivariate linear regressions on patients characteristics and tested devices

Attempts before achieving the first proper inhalation (N) Univariate analysis Multivariate analysis
delta (95 % C.I.) p delta (95 % C.I.) p
 Age a 0.08 (0.05─0.10) < 0.0001 0.07 (0.05─0.09) < 0.0001
 Sex (female) −0.03 (-0.18─0.12) NS
 Disease (COPD vs asthma) 0.36 (0.21─0.51) < 0.0001
 Previous experience with DPI 0.14 (-0.02─0.30) NS
 Previous instructed to DPI 0.12 (-0.04─0.28) NS
 Device (Breezhaler vs Genuair) 1.01 (0.84─1.18) < 0.0001 1.00 (0.84─1.16) < 0.0001
 Device (Handihaler vs Genuair) 0.91 (0.73─1.10) < 0.0001 0.87 (0.69─1.05) < 0.0001
Successful rate at the first attempt (%) Univariate analysis Multivariate analysis
OR (95 % C.I.) p OR (95 % C.I.) p
 Age a 0.85 (0.81─0.89) < 0.0001 0.84 (0.80─0.88) < 0.0001
 Sex (female) 0.96 (0.71─1.3) NS
 Disease (COPD vs asthma) 0.43 (0.31─0.58) <0.0001
 Previous experience with DPI 0.59 (0.43─0.81) < 0.001 0.64 (0.45─0.92) < 0.05
 Previous instructed to DPI 0.64 (0.47─0.88) < 0.01
 Device (Breezhaler vs Genuair) 0.17 (0.12─0.25) < 0.0001 0.15 (0.10─0.22) < 0.0001
 Device (Handihaler vs Genuair) 0.19 (0.12─0.29) < 0.0001 0.17 (0.11─0.27) < 0.0001
Total time required for the first proper inhalationb (min) Univariate analysis Multivariate analysis
delta (95 % C.I.) p delta (95 % C.I.) p
 Age a 0.24 (0.14─0.33) < 0.0001 0.25 (0.18─0.32) < 0.0001
 Sex (female) −0.29 (-0.99─0.40) NS
 Disease (COPD vs asthma) 1.12 (0.42─1.81) <0.005
 Previous experience with DPI 0.26 (-0.47─0.99) NS
 Previous instructed to DPI 0.17 (-0.55─0.88) NS
 Device (Breezhaler vs Genuair) 7.76 (7.15─8.37) < 0.0001 7.74 (7.15─8.33) < 0.0001
 Device (Handihaler vs Genuair) 2.45 (1.78─3.12) < 0.0001 2.30 (1.65─2.95) < 0.0001

DPI dry powder inhaler, NS not significant

aEvery 5-years increment

bincluding patient time and nurse demonstration