Robustness of method.

Chromatographic Conditions Variation Retention time (Minutes)
PCM MP PP CTZ
Temprature (ºC) 23 3.42 4.64 6.63 7.609
27 3.57 4.69 6.8 7.701
Flow rate (mL min-1) 0.8 3.75 4.56 6.48 7.31
1.2 3.28 4.97 6.96 7.93
Vol. of Acetonitrile (%) 43 3.67 4.71 6.75 7.83
47 3.37 4.54 6.56 7.55