Table 1.

The comparison of values between IOLMaster 500 and IOLMaster 700

n Device Mean SD Paired samples t test (p value) Range (minimum–maximum) ICC Differences of the means SD of difference of means 95% CI of differences of means Paired samples t test p value
Anterior Chamber depth (mm) 175 IOLmaster 500 3.31 0.38 <0.001 2.23–4.19 0.920 −0.08 0.13 −0.102 to −0.063 <0.001
IOLMaster 700 3.23 0.37 2.33–4.14
Axial length (mm) 171 IOLMaster 500 23.627 1.25 0.008 20.92–28.33 1.0 −0.005 0.02 −0.008 to −0.002 0.830
IOLMaster 700 23.622 1.32 20.91–29.69
Keratometry (flat, D) 188 IOLMaster 500 43.46 1.77 0.001 48.35–37.17 0.992 −0.05 0.22 −0.079 to −0.014 <0.001
IOLMaster 700 43.41 1.73 37.25–48.21
Keratometry (steep, D) 188 IOLMaster 500 44.50 1.81 <0.001 49.13–38.14 0.989 −0.08 0.27 −0.114 to −0.036 <0.001
IOLMaster 700 44.42 1.83 38.05–49.07

Difference of the means was computed by subtracting IOLMaster 500 values form IOLMaster 700 values.

ICC, intraclass correlation coefficient.