Skip to main content
. 2015 Dec 16;100(9):1201–1205. doi: 10.1136/bjophthalmol-2015-307779

Table 2.

The differences in measurements between IOLMaster 500 and IOLMaster 700; n (%)

Keratometry measurements (n=188)
Anterior chamber depth (n=175) Axial length (n=188) K1 K2
Depth (mm) n (%) Length (mm) n (%) Keratometry measurements (D) n (%) Keratometry measurements (D) n (%)
<0.1 mm 90 (52) No difference 43 (24) <0.5D 181 (96) <0.5D 173 (92)
<0.2 mm 154 (88) <0.02 mm 102 (56) <1D 188 (100) <1D 188 (100)
<0.3 mm 175 (100) <0.03 mm 150 (82)
<0.05 mm 165 (91)
<0.10 mm 177 (96)
<0.15 mm 188 (100)