Table 2: Technical Characteristics of the Reveal XT and Newer-generation Reveal LINQ Implantable Loop Recorders. The Reveal LINQ is Smaller and More Suitable for Paediatric Patients.

Characteristics Reveal LINQ Reveal XT
Q5 graphic file with name aer-05-136-g003.jpg graphic file with name aer-05-136-g004.jpg
Dimensions (mm) 44.8 × 7.2 × 4.0 62.0 × 19.0 × 8.0
Weight (g) 2.5 15
Volume (mL) 1.2 9
Duration (years) 3 3
Implantation method Injection (minimally invasive skin incision and insertion) Subcutaneous pocket
MRI compatibility 1.5 and 3 T 1.5 and 3 T
Scan time limits (minutes) No limits 30
Wireless Yes No
Algorithm for R-wave sensing Dynamic sensing threshold Dynamic sensing threshold
AF algorithm based also on P wave Yes No
Conditions detected Asystole, bradycardia, V T, V F, AT, AF Asystole, bradycardia, VT, VF, AT, AF
Automatic activation Yes Yes
Patient-activated Yes Yes
Total ECG storage (minutes) 59 (27 automatic + 2 AF + 30 manual activation) 49.5
Total ECG storage upon patient activation 30 22.5
(minutes)
Maximal length of registration per 14 pre- + 1 post-activation 14 pre- + 1 post-activation
symptomatic episode (minutes) GSM monitor Yes (at home) No
Remote transmission of data Complete Complete
Possibility of programming alarms Yes No

AF: atrial fibrillation; AT: atrial tachycardia; MRI: magnetic resonance imaging; VF: ventricular fibrillation; VT: ventricular tachycardia.