Table 2: Technical Characteristics of the Reveal XT and Newer-generation Reveal LINQ Implantable Loop Recorders. The Reveal LINQ is Smaller and More Suitable for Paediatric Patients.
Characteristics | Reveal LINQ | Reveal XT |
Q5 | ![]() |
![]() |
Dimensions (mm) | 44.8 × 7.2 × 4.0 | 62.0 × 19.0 × 8.0 |
Weight (g) | 2.5 | 15 |
Volume (mL) | 1.2 | 9 |
Duration (years) | 3 | 3 |
Implantation method | Injection (minimally invasive skin incision and insertion) | Subcutaneous pocket |
MRI compatibility | 1.5 and 3 T | 1.5 and 3 T |
Scan time limits (minutes) | No limits | 30 |
Wireless | Yes | No |
Algorithm for R-wave sensing | Dynamic sensing threshold | Dynamic sensing threshold |
AF algorithm based also on P wave | Yes | No |
Conditions detected | Asystole, bradycardia, V T, V F, AT, AF | Asystole, bradycardia, VT, VF, AT, AF |
Automatic activation | Yes | Yes |
Patient-activated | Yes | Yes |
Total ECG storage (minutes) | 59 (27 automatic + 2 AF + 30 manual activation) | 49.5 |
Total ECG storage upon patient activation | 30 | 22.5 |
(minutes) | ||
Maximal length of registration per | 14 pre- + 1 post-activation | 14 pre- + 1 post-activation |
symptomatic episode (minutes) GSM monitor | Yes (at home) | No |
Remote transmission of data | Complete | Complete |
Possibility of programming alarms | Yes | No |
AF: atrial fibrillation; AT: atrial tachycardia; MRI: magnetic resonance imaging; VF: ventricular fibrillation; VT: ventricular tachycardia.