Abstract
Objective
Hypoxaemia is a strong predictor of mortality in children. Early detection of deteriorating condition is vital to timely intervention. We hypothesise that measures of pulse oximetry dynamics may identify children requiring hospitalisation. Our aim was to develop a predictive tool using only objective data derived from pulse oximetry and observed respiratory rate to identify children at increased risk of hospital admission.
Setting
Tertiary-level hospital emergency department in Bangladesh.
Participants
Children under 5 years (n=3374) presenting at the facility (October 2012–April 2013) without documented chronic diseases were recruited. 1-minute segments of pulse oximetry (photoplethysmogram (PPG), blood oxygen saturation (SpO2) and heart rate (HR)) and respiratory rate were collected with a mobile app.
Primary outcome
The need for hospitalisation based on expert physician review and follow-up.
Methods
Pulse rate variability (PRV) using pulse peak intervals of the PPG signal and features extracted from the SpO2 signal, all derived from pulse oximetry recordings, were studied. A univariate age-adjusted logistic regression was applied to evaluate differences between admitted and non-admitted children. A multivariate logistic regression model was developed using a stepwise selection of predictors and was internally validated using bootstrapping.
Results
Children admitted to hospital showed significantly (p<0.01) decreased PRV and higher SpO2 variability compared to non-admitted children. The strongest predictors of hospitalisation were reduced PRV-power in the low frequency band (OR associated with a 0.01 unit increase, 0.93; 95% CI 0.89 to 0.98), greater time spent below an SpO2 of 98% and 94% (OR associated with 10 s increase, 1.4; 95% CI 1.3 to 1.4 and 1.5; 95% CI 1.4 to 1.6, respectively), high respiratory rate, high HR, low SpO2, young age and male sex. These variables provided a bootstrap-corrected AUC of the receiver operating characteristic of 0.76.
Conclusions
Objective measurements, easily obtained using a mobile device in low-resource settings, can predict the need for hospitalisation. External validation will be required before clinical adoption.
Keywords: Paediatric infectious disease, Pulse oximetry, Heart rate variability, Global health, Mobile health
Strengths and limitations of this study.
Features characterising pulse rate variability and blood oxygen saturation dynamics automatically derived from pulse oximetry were significantly different in children admitted to the hospital compared to those who were not admitted.
The proposed predictive model uses objective information to provide a risk score, which will allow community healthcare workers with minimal training to identify a child who may require hospital admission.
Adding pulse oximetry derived objective information to the proposed predictive model improved the identification of children at higher risk and provided improved calibration.
The proposed model has been internally validated using a bootstrap method but will require external validation.
Several pulse oximetry recordings (39%) were excluded from the study for low signal quality; thus, the app's user interface will be modified to coerce the user to improve the quality of the data.
Introduction
‘Sepsis’, a syndrome of physiological, pathological and biochemical abnormalities induced by infection, is a major public health concern and is the leading cause of death in children worldwide.1 In 2015 ∼5.9 million children under 5 years died from sepsis caused by conditions such as malaria, pneumonia, diarrhoea and measles.1 Sepsis is associated with several factors such as poverty, malnutrition, poor hygiene, low immunisation rates, poor access to care and under-resourced health systems,2 and hence its burden of mortality and morbidity is substantially higher in low and middle income countries than in the developed world.3
This disproportionate burden of mortality and morbidity from critical illnesses in children in the developing world can be prevented if early accurate diagnosis is followed by prompt treatment.3 4 Targeted interventions are needed if the sustainable development goal of 25 or fewer under-5 deaths per 1000 live births by 2030—down from 43/1000 mortality rate in 2015—is to be achieved.5
A key method to achieve this goal is to improve community case management of critical illnesses.6 In low-resource settings, this can be implemented by enabling community healthcare workers to rapidly identify critically ill children in their community allowing earlier treatment, timely referral and, ultimately, greater survival. Timely and appropriate referral will optimise use of the limited resources and can be greatly facilitated by an objective test, which identifies these children but requires a minimal amount of training or expertise.
A common feature of most treatable childhood illnesses (CI) is hypoxaemia, which correlates with disease severity and is a recognised risk factor for death.7 Including a routine and systematic approach of case finding and treating hypoxaemia is associated with improved quality of care and reduced mortality.7 8 Pulse oximetry, a simple and non-invasive method of measuring blood oxygen saturation (SpO2) and blood volume changes in tissue using the photoplethysmographic signal (PPG), is critical for the accurate detection of hypoxaemia. Thus, the use of pulse oximetry is recommended in the updated integrated management of CI (IMCI) guidelines for low and middle income countries.9 10 In addition, a reduction in heart rate variability (HRV) has been suggested as a possible early indicator of critical illness.11 HRV, traditionally calculated from the ECG signal, can be estimated using the PPG signal through a surrogate measure known as pulse rate variability (PRV).12–14 HRV and PRV have been documented to be altered in the presence of sepsis and correlated with its severity.15
Smartphones have been adopted widely, even in low-resource settings, offering an ideal platform for vital signs assessment and automatic diagnosis.16 The phone oximeter (a pulse oximeter that interfaces a finger sensor with a smartphone), for example, provides a useful tool to record pulse oximetry in low-resource settings. In a previous study, we developed a hospital admission prediction model for children aged under 5 years using clinical predictors of infectious disease and including the SpO2 recorded by the phone oximeter.17 In low-resource settings, the lack of clinical expertise for accurate identification of the signs and symptoms of infection in children is a significant roadblock to reducing mortality.4 Therefore, in this study, we aim to extend our previous work, using the same data set, by developing a tool to predict hospital admission that exclusively uses objective information, including features derived from the analysis of pulse oximetry recordings. We consider the need for hospital admission as an early indicator of critical illness, and thus this model is intended to support community healthcare workers, with limited formal training, to recognise critical illness in children earlier in the course of their disease.
Materials and methods
In order to investigate the viability of developing a hospital admission prediction test, we designed a mobile app to collect data. We then conducted a prospective observational study at the Kumudini Women's Medical College Hospital (KWMCH), a not-for-profit private tertiary-level hospital, located in the Mirzapur subdistrict of Tangail district in Bangladesh. This hospital provides primary level care to a catchment population of ∼500 000 people.
Mobile app design
We designed a customised mobile data collection app, called PhoneOxR2 (figure 1), that uses the phone oximeter for the measurement of SpO2 and HR values (figure 1A) at 1 Hz and PPG signal at 75 Hz.18 19 It also includes a respiratory rate counter interface that registers screen taps (for each inhalation) and calculates the respiratory rate after five taps.20 The hardware of the phone oximeter used in the study was an Xpod OEM pulse oximeter (Nonin Medical, Plymouth, Minnesota, USA; model 3012LP) connected to a fourth-generation iPod touch (Apple, Cupertino, California, USA). The interface of the PhoneOxR2 app displayed the PPG signal with a colour-coded (eight shades of green to red) background, which represents the real-time quality of the signal being recorded. This live quality assessment is based on the amplitude and regularity of the waveform and pulse oximeter module generated alerts.
Data collection
All individuals were recruited according to a protocol approved by the research review committee and ethical review committee of the international centre for diarrhoeal disease research, Bangladesh and the University of British Columbia/children's and women's health centre research ethics board. A research staff member explained the study and obtained written informed consent from the caretaker accompanying the child.
Children with chronic diseases or with previously documented low SpO2 levels due to other conditions, such as cardiac disease, were excluded.17 All children were assessed and managed by a team of physicians not involved in the study. Pulse oximetry information was blinded (not displayed) within the app (figure 1B) and not provided to the physicians, as this was not routinely available in the facility. Thus, the SpO2 recording did not influence clinical decisions, including the need for admission, which was decided at the physicians' discretion uninfluenced by the study. The children were considered admitted under the following circumstances: (1) admitted and stayed for more than 24 hours in the hospital, (2) sent home initially but admitted at a subsequent visit within 1 week, (3) were advised for admission but the advice was not followed and (4) were transferred or referred to another facility providing a higher level of care. The most common diagnoses included acute lower or upper respiratory infection, and eye or ear infection in admitted children; and acute upper respiratory infection, eye or ear infection, diarrhoea, fever of unknown cause and gastrointestinal symptoms in non-admitted children; see ref. 17 for more details of the distribution of diagnoses and outcomes. From the children sent home, we were able to follow-up 1855/2514 (73.8%) and 14 were admitted on a subsequent visit.
The data acquisition was performed, using the PhoneOxR2, between October 2012 and April 2013 (winter is from November to February). We collected 1 min recordings of the SpO2 and PPG signals along with respiratory rate from a total of 3374 children under 5 years presenting at the outpatient or emergency department. The respiratory rate was estimated from the PPG signal through a previously proposed algorithm21 in 39 cases when the respiratory tapping interface was not used.
Pulse oximetry processing
The quality of each recorded PPG signal was evaluated offline, using an algorithm that assigns a signal quality index (SQI) between 0 and 100 (100 being the best quality). The method is based on pulse segmentation and cross-correlation of consecutive pulse segments.22 PPG segments with an SQI higher than 50 were automatically identified for further analysis. The analysis was limited to children for whom there was at least 30 s of good quality (SQI>50%) PPG signal. PRV and SpO2 derived features were computed for each data recording.
PRV characterisation
We applied a peak detection algorithm based on zero-crossing to locate the pulse peaks in the PPG signal and to obtain pulse-to-pulse interval time series (PPIs).23 Five time-domain parameters were computed from the PPIs: (1) the mean of the PPIs: representing the RR interval or the time interval between two consecutive R waves in the ECG, (2) the SD of PPIs: representing SDNN or the SD of the so-called normal-to-normal (NN) intervals, (3) the root mean square of the successive differences (RMSSD) between adjacent PPIs: representing RMSSD or the RMSSD between adjacent NN intervals and (4 and 5) two standard descriptors to evaluate non-linear short-term and long-term variability (SD1 and SD2) based on the Poincaré plot of PPIs.24
For the frequency analysis, each segment of PPIs was resampled into an equivalent, uniformly spaced time series (sampling rate of 4 Hz), and the power spectral density was computed. The power in each frequency band was computed by integrating the area under the power spectral density curve bound by the band of interest: very low frequency (0.01–0.04 Hz), low frequency (LF; 0.04–0.15 Hz) and high frequency (HF; 0.15–0.4 Hz). LF and HF powers were normalised (LFn and HFn) by the total spectral power between 0.04 and 0.4 Hz. The ratio of the low-to-HF power (LF/HF ratio) was also computed.
SpO2 characterisation
Several oximetry indices proposed in previous studies to assess SpO2 dynamics were computed.25 These included the number of SpO2 desaturations >1%, 2% and 3% below baseline (nb1, nb2 and nb3, respectively), cumulative time spent below an SpO2 of 98%, 96% and 94% (tb98, tb96 and tb94, respectively) and the Delta index (SpO2delta).26 The Delta index quantifies SpO2 variability and was computed as the average of absolute differences of the mean oxygen saturation between successive 12 s intervals. The median (SpO2median) and SD (SpO2std) of the SpO2 within each segment were also calculated.
Data analysis
All data analysis was conducted using R v3.2.0 (R Foundation for statistical computing, Vienna, Austria). To facilitate interpretation, raw heart rate and respiratory rate were standardised with respect to age using population median and SD values.27
Univariate analysis
The relationship between each feature and the need for facility admission was assessed by comparing the median and IQR between the admitted and the non-admitted participants and through univariate logistic regression using the OR (95% CI). The logistic regression analysis for all PRV features was adjusted for age, a known confounding factor.28 For these features, we added age into the univariate models to compute the adjusted ORs.
Multivariate model development
A stepwise selection method was applied to select the most relevant features and to develop the final multivariate logistic regression model, which we refer to as the ‘mobile model’. The stepwise selection method added/dropped one feature at a time and stopped when further inclusion/exclusion no longer improved the model, as determined by the Akaike information criterion.29 All features were included as candidate predictors and a minimum value of 40 events per variable was maintained to avoid overfitting.30 The stepwise selection method was also compared with the least absolute shrinkage and selection operator (LASSO) method (computed via the glmnet R package31). The calibration of the model was evaluated by the Hosmer-Lemeshow goodness-of-fit test. To assess the impact of including objective information derived from pulse oximeter recordings, the performance of the ‘mobile model’ was compared to the ‘baseline-model’, a logistic regression model including only respiratory rate and the median SpO2 value.
Model validation
The model was validated using the bootstrap method, recommended as the optimal technique for estimation of internal validity of a predictive logistic regression model.30 Using the original data set, 100 bootstrap samples were generated using sampling with replacement. The stepwise selection procedure was applied to each bootstrap sample to develop a multivariate model. This model was applied to predict the outcome in the bootstrap sample and in the original data set. The difference in the prediction performance, quantified by the area under the receiver operating characteristic curve (AUC), was computed to estimate the optimism of the ‘mobile model’.
Risk threshold determination
The risk threshold for the estimated probability of facility admission was chosen to maximise a weighted classification score defined as (the number of correct identifications of true positives)+(the number of correct identifications of true negatives). The weighted classification score was computed for various previously established ratios of false negative cases to false positive cases (1:3, 1:5 and 1:10);17 however, the ratio should be predetermined in each specific clinical context.
Model classification performance
Once the risk threshold was set, the performance of the ‘mobile model’ was evaluated in terms of accuracy, sensitivity, positive predictive value (PPV) and negative predictive value (NPV) and compared to the ‘baseline model’. The reclassification technique, offering incremental information over the AUC, was employed to compare both models using the net reclassification improvement (NRI) and integrated discrimination improvement (IDI).32 The NRI quantified the performance of the ‘mobile model’ reclassifying children as compared to the ‘baseline model’. The IDI measured the improvement of the ‘mobile model’ relative to the ‘baseline model’ and was quantified as the amount of increased difference in the average predicted risk between admitted or non-admitted children.32
Results
Excluded cases
In total, 2051 (61%) of the 3374 recruited children had at least 30 s of good quality PPG signal recorded, which permitted the adequate characterisation of SpO2 and PRV. The remaining 1323 children were excluded from analysis. The percentage of excluded cases was similar in the admitted (40%) and non-admitted (38%) groups. The SpO2 values of the excluded cases were not significantly different from those included, evaluated through univariate logistic regression. Thus, there was no clear association between the patient's risk and the PPG signal quality.
Univariate analysis
Pulse oximetry derived features were significantly different in children who required admission compared to those who did not require admission (table 1). Admitted children presented a significantly higher respiratory rate, higher HR and lower PRV (reflected through reduced SDNN, RMSSD, SD1 and SD2 values) than non-admitted children. Admitted children also showed reduced PRV in the spectral domain suggested by lower power in low (LFn) and high (HFn) frequency bands, relative to non-admitted children (table 1).
Table 1.
Feature | Admission required N=616 (30.0%) | Admission not required N=1435 (70.0%) | p Value | OR (95% CI) |
---|---|---|---|---|
Pulse rate variability analysis* | ||||
Heart rate (bpm) | 140 (126–154) | 125 (112–138) | <0.001 | 1.03 (1.02 to 1.03) |
RR (s) | 0.43 (0.39–0.47) | 0.48 (0.43–0.53) | <0.001 | 0.49 (0.42 to 0.57)† |
SDNN (s) | 0.016 (0.011–0.025) | 0.022 (0.014–0.037) | <0.001 | 0.83 (0.78 to 0.89)‡ |
RMSSD (s) | 0.016 (0.013–0.025) | 0.021 (0.015–0.039) | <0.001 | 0.91 (0.87 to 0.95)‡ |
SD1 (s) | 0.012 (0.093–0.018) | 0.015 (0.011–0.027) | <0.001 | 0.99 (0.98 to 0.99)‡ |
SD2 (s) | 0.018 (0.011–0.029) | 0.026 (0.016–0.043) | <0.001 | 0.98 (0.98 to 0.99)‡ |
LFn (nu) | 0.37 (0.2–0.57) | 0.38 (0.23–0.55) | 0.003 | 0.93 (0.89 to 0.98)‡ |
HFn (nu) | 0.14 (0.09–0.24) | 0.18 (0.12–0.27) | <0.001 | 0.86 (0.79 to 0.94) |
LF/HF (ratio) | 2.3 (1.1–4.7) | 2 (0.95–3.9) | 0.2 | 1.00 (0.99 to 1.00) |
SpO2 analysis | ||||
SpO2median (%) | 96 (91–98) | 98 (97–99) | <0.001 | 0.79 (0.76 to 0.81) |
SpO2std (%) | 0.73 (0.51–1.2) | 0.54 (0.4–0.83) | <0.001 | 0.54 (0.4 to 0.83) |
SpO2delta (%) | 0.31 (0.19–0.48) | 0.22 (0.13–0.34) | <0.001 | 3.5 (2.5 to 4.8) |
tb98 (s) | 39 (11–56) | 3.3 (0–34) | <0.001 | 1.4 (1.3 to 1.4)§ |
tb96 (s) | 26 (0–53) | 0 (0–3.2) | <0.001 | 1.4 (1.3 to 1.5)§ |
tb94 (s) | 2 (0–41) | 0 (0–0.33) | <0.001 | 1.5 (1.4 to 1.6)§ |
nb1 (# times) | 4 (1–8) | 1 (0–4.5) | <0.001 | 1.1 (1.08 to 1.12) |
nb2 (# times) | 0 (0–3) | 0 (0–1) | <0.001 | 1.1 (1.1 to 1.2) |
nb3 (# times) | 0 (0–1) | 0 (0–0) | <0.001 | 1.1 (1.1 to 1.2) |
Non-pulse oximetry data | ||||
Heart rate age z-score | 0.66 (−0.19 to 1.7) | 0.23 (−0.58 to 1.1) | <0.001 | 1.3 (1.2 to 1.4) |
Respiratory rate age z-score | 0.47 (−0.48 to 1.9) | −0.046 (−0.8 to 1.1) | <0.001 | 1.2 (1.1 to 1.3) |
Age (days) | 230 (70.5–542) | 477 (202–1038) | <0.001 | 0.989 (0.987 to 0.991)§ |
Gender (% of male) | 67% | 59% | <0.001 | 1.43 (1.17 to 1.74) (male vs female) |
Values include feature medians (quartiles), p value and OR (95% CI).
*Adjusted for age.
†Associated with an increase of 0.1 times the feature unit.
‡Associated with an increase of 0.01 times the feature unit.
§Associated with an increase of 10 times the feature unit.
HFn, high frequency normalised; LFn, low frequency normalised; RMSSD, root mean square of the successive differences; RR, respiratory rate; SpO2, blood oxygen saturation.
The SpO2 (median) was lower and showed more variability reflected via several features (SpO2std, delta, tb98, tb96, tb94, nb1, nb2, nb3) in admitted compared to non-admitted children. Children who required admission were also significantly younger and more likely to be male (table 1).
Final model validation
The majority of features selected by the stepwise method were also selected by the LASSO selection method. For simplicity, we decided to use the feature selection result from the stepwise method to construct the ‘mobile model’, which includes only information derived from mobile device recordings. All the selected features had p<0.05 and the ‘mobile model’ equation was: logit (probability (admission required))=3.806–0.049×(SpO2 median)+0.012×(time spent below an SpO2 of 98%)+0.018×(time spent below an SpO2 of 94%)+0.152×(respiratory rate age z-score)−0.033×(age in month)+0.200×(heart rate age z-score)+0.280×(male=1, female=0)−0.656(normalised power in the low frequency (LF) band). This model presented an AUC of 0.77 (figure 2A) and was well calibrated (Hosmer-Lemeshow goodness of fit test p=0.84).
For comparison purposes, the implemented ‘baseline model’ equation was: logit(probability (admission required))=21 461+0.090×(respiratory rate age z-score)−0.233×(SpO2 median). This model presented an AUC of 0.729 (figure 2B).
With regard to the internal validation, each of the eight final selected predictors was also selected in at least 70% of the bootstrap samples indicating a stable predictive capability. The ‘mobile model’ had optimism in AUC 0.012 (95% CI −0.01 to 0.03) and the bootstrap-corrected AUC was 0.754 (95% CI 0.731 to 0.778).
Model classification
The maximum points indicated in figure 3 illustrate the risk thresholds that maximised the weighted classification score based on the studied false negative to false positive ratios.
Consider a context in which the cost of sending healthy children to hospital is not overwhelmingly high, for example, in a community close to a major hospital. It would be reasonable to allow an unnecessary admission of three healthy children (ie, three false positives) in order to avoid missing any seriously sick child (ie, one false negative). As illustrated in figure 3, the optimal risk threshold in this context (3:1) was 0.265 for the ‘mobile model’ and 0.26 for the ‘baseline model’. The classification results were obtained with an adjusted risk threshold of 0.25 for easier interpretation (table 2). The ‘mobile model’ achieved better classification results in terms of NPV and sensitivity. However, both methods provided a low PPV (table 2).
Table 2.
Model | Risk threshold | PPV (95% CI) | NPV (95% CI) | Sensitivity (95% CI) | Specificity (95% CI) |
---|---|---|---|---|---|
‘Mobile’ | 0.25 | 0.51 (0.48 to 0.55) | 0.85 (0.83 to 0.87) | 0.72 (0.68 to 0.75) | 0.71 (0.69 to 0.73) |
‘Baseline’ | 0.25 | 0.51 (0.48 to 0.55) | 0.82 (0.80 to 0.84) | 0.63 (0.59 to 0.67) | 0.74 (0.72 to 0.77) |
NPV, negative predictive value; PPV, positive predictive value.
Reclassification results for both models showed that the addition of objective information derived from pulse oximetry recordings (SpO2 dynamics and PRV analysis) exhibited a 28% net improvement in the classification of the admitted (figure 4) and 4% net improvement in the classification of the non-admitted patients (figure 4), at a risk threshold of 0.25. The IDI was 0.0466 (95% CI 0.0368 to 0.0563, p=0) and the categorical NRI was 0.0389 (95% CI 0.0042 to 0.0736, p=0.02799).
Discussion
We have developed and internally validated a prediction model that uses objective information to identify under-5 children who would require hospitalisation. The objective information includes age, gender, respiratory rate and features derived from the analysis of 1 min pulse oximetry recordings. In this study, we have shown that HRV estimated using the PPG signal (PRV), SpO2 dynamics and respiratory rate are significantly different in children admitted to the hospital. A high respiratory rate and HR, suppressed PRV-power in the LF band, low median SpO2 value and high SpO2 variability, together with age and gender information, were the strongest predictors of the need for hospitalisation. When a recording with adequate signal quality (>50%) was applied, the model using these features provided a bootstrap-corrected AUC of 0.754. This enhanced-model provided a small but significant improvement in classification performance relative to a model using respiratory rate and median SpO2 value alone. This improvement is most noticeable as an enhanced ability to identify children at risk with only a small increase in the number of false positive results.
In low-resource settings, there is a lack of healthcare workers who are adequately trained for early detection of critical illness in children. One of the advantages of the proposed predictive model is that it exclusively uses objective information extracted primarily from pulse oximetry recordings (PRV and SpO2 dynamics) to generate a risk score. This information can be automatically calculated by a mobile application without requiring any additional effort as long as a good quality pulse oximetry signal is obtained. Thus, community healthcare workers can easily use it without requiring specific additional training for recognising signs or symptoms associated with critical illness. Providing community healthcare workers with this mobile–health approach will facilitate a rapid detection of children that would require admission to a facility and has the potential to modify behaviour towards strengthening emergency response systems and improving decision support information availability at a community level. When implemented within the phone oximeter, this could offer a convenient and easy-to-use platform for improving hospital referral practices in low-resource settings. The current version of the phone oximeter uses a much lower cost finger sensor directly connected to a smartphone or tablet through the audio jack.33
The selection of the risk threshold for predicted probability is a trade-off between acceptable risk and undue burden to the family, patient and health system and should be predetermined according to the clinical context. At a risk threshold of 0.25, the model provided a high-NPV (85%), demonstrating that the model is able to accurately identify children who do not require referral to a facility, without neglecting to refer many patients who should receive further attention. However, it provided a low PPV, suggesting that only 51% of children referred to the hospital would require admission. Increasing sensitivity or negative predictive performance will be at the expense of increasing unnecessary referral for hospital admission. The range of risk could also be used to individualise the frequency of follow-up or to follow the progression of risk overtime.
In a previous study, we developed a clinical prediction model based on continuous oxygen saturation, respiratory rate, temperature and age, combined with additional subjective predictors such as lethargy, irritability, symptoms of cough, diarrhoea and fast or difficult breathing, to predict the need for hospitalisation in children.17 Although this model provided improved performance (AUC=0.89), the ability to accurately identify some of the signs and symptoms may be difficult in settings where frontline healthcare workers are inadequately trained in the early detection of these disease signs and symptoms—a common scenario in low-resource settings. Therefore, in this study, we aimed to provide a predictive model based only on objective measures that can be obtained with minimal specific training.
Our findings from PPG analysis suggest that children with suppressed PRV, both in the time and frequency domains, and increased heart rate were more likely to be admitted. Children at a higher risk of critical illness have been previously demonstrated to have a higher heart rate and lower variability between ECG-based heart beats as measured in the time domain.11 Time-domain features based on heart rate characteristics have also been used in the early diagnosis of neonatal sepsis providing an AUC of 0.73.34 35 Spectral HRV and PRV analysis has also been used to identify patients with severe sepsis in the emergency department12 28 and sepsis shock in intensive care.36 These studies highlighted the value of spectral analysis for monitoring the progression of sepsis. The power within the LF band, which reflects sympathetic modulation of heart rate, was demonstrated to be negatively correlated with sepsis severity and the risk of sepsis shock.12 28 36 This feature was also selected in the current study as the strongest PRV-based risk predictor, confirming the presence of abnormal autonomic control in admitted children. The analysis of the SpO2 signal, recorded once per second, allowed us to assess not only the SpO2 levels but also the short-term SpO2 dynamics, not previously included as sepsis predictors.
The addition of these significant features, extracted from the analysis of PRV and SpO2 dynamics to the prediction model using only the SpO2 median and respiratory rate, improved the classification by 28% in admitted children and by 4% in non-admitted children. Thus, the information derived from pulse oximetry has the potential to provide relevant objective information that improves the identification of children at higher risk. The updated IMCI guidelines recommend the use of pulse oximetry to identify hypoxaemia and improve the diagnosis of critical illness. Using the phone oximeter, pulse oximetry can provide even more objective information, the derived PRV and SpO2 dynamics, without increasing the required effort. The app will automatically calculate this information, unbiased by the community healthcare worker. The result of the app can then be considered together with treatment guidelines for a more informed decision on the level of care necessary for the child.
Limitations and future research
The most significant limitation of this study is that 39% of the cases had to be excluded from the study because they did not meet the signal quality criteria for further analysis. The interface in the PhoneOxR2 software was designed to assist the user in optimising data collection. The colour of the background reflected the current quality of the signal, estimated using amplitude and regularity of the PPG signal and pulse oximeter module generated alerts. Feedback on the overall quality was provided at the conclusion of the recording and users were also trained to optimise the signal quality during the 1 min recording. However, the signal-processing methods applied in this study require a high-quality PPG signal and thus demand a more conservative SQI, such as the one applied offline in this study.26 We have subsequently implemented this more restrictive SQI method in the new app and made significant improvements in the design of new sensors. Any future decision support app that includes the predictive model will include this updated SQI version.26 Additionally, the app will require at least 30 s of good quality data (SQI>50%) to be accepted as a successful recording, a restriction that was not enforced during data collection.
The primary outcome, the need for hospitalisation, was a subjective decision made by the facility physicians, who may have been overly conservative in their decision-making, knowing that their clinical outcomes were being studied. To reduce this possible bias, the cases that were discharged within 24 hours were excluded and all pulse oximetry data were not provided to the facility physicians. Nevertheless, this study showed that there are significant differences in pulse oximetry derived features between admitted and non-admitted children and that this measurable objective information can correlate with a subjective decision, such as hospital admission.
The final model was internally validated through bootstrapping. However, external validation of the proposed predictive model should be performed in the setting of intended use. Owing to the practicality of collecting a large enough sample size and the need for good outcome data, our acquisition was carried out at a tertiary-level facility rather than in the community. The model should be validated in a community setting, with data from a cohort of children assessed by community healthcare providers. The model and thresholds for referral will need to be optimised for the local context.
Another limitation is the modest values of PPV and specificity. While this performance is significantly better than the currently used criteria (such as respiratory rate and SpO2 alone), there are a number of options for improving the model performance. We are investigating the specific characteristics of those participants in the current cohort who were misclassified. We are also looking to identify additional objective information that could be included to improve performance. One option is to add the morphology of the PPG waveform, which also provides information about the cardiovascular system and has been previously used as a measure of fluid responsiveness.37 38
Conclusion
Following further validation and improved quality of signal capture, this prediction tool could be used to improve the health outcomes of children under 5 years in low and middle income countries. The benefit of this predictive model is that it only uses objective information easily recorded from a mobile device, to provide a risk score for the probability of the need for the child to be admitted to a facility. This could allow frontline health workers with a mobile device to rapidly detect the need for referral with limited training overhead.
Acknowledgments
The authors would like to thank the clinical staff of the Kumudini Women's Medical College Hospital in Bangladesh for their collaboration and assistance with data acquisition, and the study coordinators from the University of British Columbia and British Columbia Children's Hospital who contributed to the study design and analysis.
Footnotes
Contributors: All authors contributed to the design of the study, and in critically revising and finally approving the manuscript. SR, TH and SEA performed the experiments. AG and PD performed the signal processing and extracted the objective features. AG, GZ, DD and JMA analysed the data. AG, GZ, DD, WK, PD, GAD and JMA contributed with materials/analysis tools. AG, GZ, SR, DD, WK, PD, TH, SEA, CL, NK, GAD and JMA wrote the manuscript.
Funding: This study was a component of the project titled Interrupting Pathways to Maternal, Newborn and Early Childhood Sepsis (IPSI) (Project # S065353–001) that was funded by the Canadian Department of Foreign Affairs, Trade and Development's MUSKOKA Initiative on Maternal, Newborn and Child Health (MNCH), received by CL, http://www.international.gc.ca/.
Competing interests: JMA and GAD are founders of LionsGate Technologies (LGTmedical) that produces low-cost mobile device-based pulse oximetry devices.
Patient consent: Obtained.
Ethics approval: Institutional review board approval was obtained from the Research Review Committee (RRC) and Ethical Review Committee (ERC) of the Institutional Review Board of International Centre for Diarrhoeal Disease Research, Bangladesh and the University of British Columbia/Children's and Women's Health Centre Research Ethics Board.
Provenance and peer review: Not commissioned; externally peer reviewed.
Data sharing statement: No additional data are available.
References
- 1.You D, Hug L, Ejdemyr S et al. . Global, regional, and national levels and trends in under-5 mortality between 1990 and 2015, with scenario-based projections to 2030: a systematic analysis by the UN Inter-agency Group for Child Mortality Estimation. Lancet 2015;6736:1–12. [DOI] [PubMed] [Google Scholar]
- 2.UNICEF. Committing to Child Survival: A Promise Renewed. 2012. UNICEF Progress Report, 2012. http://www.un.org/ru/publications/pdfs/renewed_%20progress%20report%202012.pdfhttp://www.un.org/ru/publications/pdfs/renewed_%20progress%20report%202012.pdf [Google Scholar]
- 3.Khilanani A, Mazwi M, Paquette ET. Pediatric sepsis in the global setting. Clin Pediatr Emerg Med 2014;15:193–203. 10.1016/j.cpem.2014.04.006 [DOI] [Google Scholar]
- 4.Mangia CMF, Kissoon N, Carcillo JA. Sepsis and septic shock: a global overview. J Pediatr Infect Dis 2009;4:71–6. [Google Scholar]
- 5.Liu L, Johnson H, Cousens S. Global, regional, and national causes of child mortality in 2000–2010: an updated systematic analysis. Lancet 2015;385:430–40. 10.1016/S0140-6736(14)61698-6 [DOI] [PubMed] [Google Scholar]
- 6.Theodoratou E, Al-Jilaihawi S, Woodward F et al. . The effect of case management on childhood pneumonia mortality in developing countries. Int J Epidemiol 2010;39:155–71. 10.1093/ije/dyq032 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7.Subhi R, Adamson M, Campbell H et al. . The prevalence of hypoxaemia among ill children in developing countries: a systematic review. Lancet Infect Dis 2009;9:219–27. 10.1016/S1473-3099(09)70071-4 [DOI] [PubMed] [Google Scholar]
- 8.Lazzerini M, Sonego M, Pellegrin MC. Hypoxaemia as a mortality risk factor in acute lower respiratory infections in children in low and middle-income countries: systematic review and meta-analysis. PLoS ONE 2015;10:e0136166 10.1371/journal.pone.0136166 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.World Health Organization. Pocket book of hospital care for children: guidelines for the management of common childhood illnesses – 2nd ed, 2013. http://apps.who.int/iris/bitstream/10665/81170/1/9789241548373_eng.pdfhttp://apps.who.int/iris/bitstream/10665/81170/1/9789241548373_eng.pdf [PubMed]
- 10. World Health Organization. Recommendations for management of common childhood conditions, 2012. http://apps.who.int/iris/bitstream/10665/44774/1/9789241502825_eng.pdf.
- 11.Ahmad S, Tejuja A, Newman KD et al. . Clinical review: a review and analysis of heart rate variability and the diagnosis and prognosis of infection. Crit Care 2009;13:232 10.1186/cc8132 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Barnaby D, Ferrick K, Kaplan DT et al. . Heart rate variability in emergency department patients with sepsis. Acad Emerg Med 2002;9:661–70. 10.1111/j.1553-2712.2002.tb02143.x [DOI] [PubMed] [Google Scholar]
- 13.Gil E, Orini M, Bailón R et al. . Photoplethysmography pulse rate variability as a surrogate measurement of heart rate variability during non-stationary conditions. Physiol Meas 2010;31:1271–90. 10.1088/0967-3334/31/9/015 [DOI] [PubMed] [Google Scholar]
- 14.Dehkordi P, Garde A, Karlen W et al. . Pulse rate variability compared with heart rate variability in children with and without sleep disordered breathing. Conf Proc IEEE Eng Med Biol Soc 2013;2013:6563–6. [DOI] [PubMed] [Google Scholar]
- 15.Ahmad S, Ramsay T, Huebsch L et al. . Continuous multi-parameter heart rate variability analysis heralds onset of sepsis in adults. PLoS ONE 2009;4:e6642 10.1371/journal.pone.0006642 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Kahn JG, Yang JS, Kahn JS. ‘Mobile’ health needs and opportunities in developing countries. Health Aff (Millwood) 2010;2:252–8. 10.1377/hlthaff.2009.0965 [DOI] [PubMed] [Google Scholar]
- 17.Raihana S, Dunsmuir D, Huda T et al. . Development and internal validation of a predictive model including pulse oximetry for hospitalization of under-five children in Bangladesh. PLoS ONE 2015;10:e0143213 10.1371/journal.pone.0143213 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Karlen W, Dumont G, Petersen C et al. . Human-centered Phone Oximeter interface design for the operating room. In: HEALTHINF. 2011 Presented at: International Conference on Health Informatics; 2011; Rome, Italy. 433–438. 10.5220/0003335204330438 [DOI] [Google Scholar]
- 19.Dunsmuir D, Petersen C, Karlen W et al. . The Phone Oximeter for mobile spot-check. In: Anest Analg. 2012. Presented at: 2012 Society for Technology in Anesthesia Annual Meeting; January 18-21, 2012; Palm Beach, FL. 26. [Google Scholar]
- 20.Karlen W, Gan H, Chiu M et al. . Improving the accuracy and efficiency of respiratory rate measurements in children using mobile devices. PLoS ONE 2014;9:e99266 10.1371/journal.pone.0099266 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Karlen W, Garde A, Myers D et al. . Estimation of respiratory rate from photoplethysmographic imaging videos compared to pulse oximetry. IEEE J Biomed Heal Inform 2015;19:1331–8. 10.1109/JBHI.2015.2429746 [DOI] [PubMed] [Google Scholar]
- 22.Karlen W, Kobayashi K, Ansermino JM et al. . Photoplethysmogram signal quality estimation using repeated Gaussian filters and cross-correlation. Physiol Meas 2012;33:1617–29. 10.1088/0967-3334/33/10/1617 [DOI] [PubMed] [Google Scholar]
- 23.Dehkordi P, Garde A, Karlen W et al. . Evaluation of cardiac modulation in children in response to apnea / hypopnea using the phone oximeter(TM). Physiol Meas 2016;37:187–202. 10.1088/0967-3334/37/2/187 [DOI] [PubMed] [Google Scholar]
- 24.Hoshi RA, Pastre CM, Vanderlei LCM et al. . Poincaré plot indexes of heart rate variability: relationships with other nonlinear variables. Auton Neurosci 2013;177:271–4. 10.1016/j.autneu.2013.05.004 [DOI] [PubMed] [Google Scholar]
- 25.Garde A, Dehkordi P, Karlen W et al. . Development of a screening tool for sleep disordered breathing in children using the phone oximeterTM. PLoS ONE 2014;9:e112959 10.1371/journal.pone.0112959 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Levy P, Pepin JL, Paramelle B et al. . Accuracy of disturbances oximetry for detection in sleep respiratory apnea syndrome. Chest 1996;109:395–9. 10.1378/chest.109.2.395 [DOI] [PubMed] [Google Scholar]
- 27.Fleming S, Thompson M, Stevens R. Normal ranges of heart rate and respiratory rate in children from birth to 18 years of age—supplementary webappendix. Lancet 2011;377:1011–18. 10.1016/S0140-6736(10)62226-X [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Middleton PM, Tang CHH, Chan GSH et al. . Peripheral photoplethysmography variability analysis of sepsis patients. Med Biol Eng Comput 2011;49:337–47. 10.1007/s11517-010-0713-z [DOI] [PubMed] [Google Scholar]
- 29.Akaike H. Information theory and an extension of the maximum likelihood principle. In: Petrov BN, Csaki F, eds, Second international symposium on information theory. Budapest: Academiai Kiado, 1973:267–81. [Google Scholar]
- 30.Steyerberg EW, Harrell FE, Borsboom GJJ et al. . Internal validation of predictive models. J Clin Epidemiol 2001;54:774–81. 10.1016/S0895-4356(01)00341-9 [DOI] [PubMed] [Google Scholar]
- 31.Friedman J, Hastie T, Tibshirani R. Regularization paths for generalized linear models via coordinate descent. J Stat Softw 2009;30:1–3. [PMC free article] [PubMed] [Google Scholar]
- 32.Pencina MJ, D'Agostino RB Sr, D'Agostino RB Jr et al. . Evaluating the added predictive ability of a new marker: from area under the ROC curve to reclassification and beyond. Stat Med 2008;27:157–72; discussion 207–12 10.1002/sim.2929 [DOI] [PubMed] [Google Scholar]
- 33.Petersen CL, Chen TP, Ansermino JM et al. . Design and evaluation of a low-cost smartphone pulse oximeter. Sensors (Basel) 2013;13:16882–93. 10.3390/s131216882 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Griffin MP, O'Shea TM, Bissonette EA et al. . Abnormal heart rate characteristics preceding neonatal sepsis and sepsis-like illness. Pediatr Res 2003;53:920–6. 10.1203/01.PDR.0000064904.05313.D2 [DOI] [PubMed] [Google Scholar]
- 35.Griffin MP, Lake DE, Moorman JR. Heart rate characteristics and laboratory tests in neonatal sepsis. Pediatrics 2005;115:937–41. 10.1542/peds.2004-1393 [DOI] [PubMed] [Google Scholar]
- 36.Piepoli M, Garrard CS, Kontoyannis DA et al. . Autonomic control of the heart and peripheral vessels in human septic shock. Intensive Care Med 1995;21:112–19. 10.1007/BF01726532 [DOI] [PubMed] [Google Scholar]
- 37.Addison PS. A review of signal processing used in the implementation of the pulse oximetry photoplethysmographic fluid responsiveness parameter. Anesth Analg 2014;119:1293–306. 10.1213/ANE.0000000000000392 [DOI] [PubMed] [Google Scholar]
- 38.Sahni R. Noninvasive monitoring by photoplethysmography. Clin Perinatol 2012;39:573–83. 10.1016/j.clp.2012.06.012 [DOI] [PubMed] [Google Scholar]