Table 1.

Calibration and cross validation results for fluorine. C-set: calibration set; DW: Durbin–Watson; NIR: near infrared; RPDcal: relative prediction deviation of calibration; RPDCV: relative prediction deviation of cross-validation; SD: standard deviation; SEC: standard error of calibration; SECV: standard error of cross-validation; SEL: standard error of the laboratory; WN: wavenumber.

Parameter Fluorine
Units ppm
SEL-reproducibility 17.34
#samples 39
Outliers 0
Min 0.00
Mean 346.92
Max 800.00
SD 260.17
WN range/step (cm−1) 9000–4500/8
Pretreatments SNV/D1
Regression method PLS
Number of factors 4
SEC 86.14
R2cal 0.9471
SECV 76.83
R2CV 0.9464
NIR repeatability 0.20
DW 2.29
C-Set Durbin–Watson in range 1.5 to 2.5? yes
Q-value 0.85
RPDcal 2.62
RPDCV 3.60