Table 1. πSPIM resolution, anisotropy, and focal volumes.

Mode Objective(s) NA Resolution (FMWH)/nm Anisotropy Focal volume/fl
type     x,y (theory) z (theory) (theory) (theory)
widefield PlanApoW25× 1.1 274 (240) 785 (801) 0.35 (0.30) 0.072 (0.057)
πSPIM PlanApoW25× & ApoTIRF60× 1.1/1.49* 284 (212) 339 (285) 0.84 (0.74) 0.057 (0.023)
widefield PlanApoW60× 1.27 250 (203) 648 (510) 0.39 (0.40) 0.052 (0.027)

Abbreviations: NA, numerical aperture; FWHM, full-width-at-half-maximum, anisotropy = FWHMx,y/FWHMz.