Skip to main content
. 2016 Jan 23;26(10):3691–3696. doi: 10.1007/s00330-015-4183-4

Table 2.

Radiation dose in head CT: Adaptive Statistical Iterative Reconstruction (ASiR) compared to filtered back projection (FBP)

ASiR FBP ASiR vs. FBP p-value
DLP (mGy*cm)
 Overall 654.5 ± 77.0 804.2 ± 52.3 -19 % <0.0001
 Cerebrum 390.3 ± 75.5 560.6 ± 51.8 -30 % <0.0001
 Posterior fossa 264.2 ± 31.9 243.6 ± 5.5 +8 % <0.0001
CTDIvol (mGy)
 Cerebrum 36.2 ± 4.5 51.5 ± 3.0 -30 % <0.0001
 Posterior fossa 59.7 ± 3.7 55.9 ± 2.5 +7 % <0.0001
 Scan range (cm) 14.3 ± 1.1 14.6 ± 1.1 +2 % NS

DLP dose-length product, CTDI vol volumetric CT dose index