Table 4.

Comparison of radiation-dose from head CT examinations with Adaptive Statistical Iterative Reconstruction (ASiR) image reconstruction

Head CT protocols with ASiR
Scan mode CTDIvol Total DLP
Posterior fossa Brain
Our study Axial 59.2 36.4 654.5
Kilic et al. 2011 [7] Axial 93.5 38.6 748.5
Ren et al. 2012 [17] Helical 38.1 559.0
Rapalino et al. 2012 [8] Helical 49.7 932.2
Komlosi et al. 2014 [9] Helical 57.1 1,190.9
AAPM 20121[16] Axial 69.7 41.9 Not provided
Helical 38.4 Not provided

1The collection of routine head CT protocols by the American Association of Physicists in Medicine for different scanner vendors and models. Radiation doses for GE Discovery 750 HD with are ASiR provided here.

CTDIvol volume CT dose index, DLP dose length product