Table 3. Root-mean-squared-errors (RMSEs) between estimated and measured central BP [mmHg].

Central BP Estimates Low PP Amplification (1.06 ± 0.07) Middle PP Amplification (1.25 ± 0.07) High PP Amplification (1.59 ± 0.13)
TW SP PP TW SP PP TW SP PP
Radial BP 6.6* 6.1# 6.1 7.8* 13.9* 13.9* 8.1* 21.6* 21.6*
GTFSphygmoCor 4.7* 7.5* 10.1* 3.5 5.4 7.9* 2.9 3.1 4.8
GTFARX 5.2* 6.2# 7.1* 3.2 3.5 4.6 2.9 3.5 4.3
ATF 3.5 3.3 4.2 3.5 3.3 3.4 3.1 3.7 3.7

PP amplification is the ratio of radial to central pulse pressure; TW, central total waveform; SP, central systolic pressure; and PP, central pulse pressure. *and # denote statistically different (e.g., p < 0.05) or borderline statistically different (e.g., p ≈ 0.05) compared to ATF, respectively.