Skip to main content
. 2015 Nov 3;1(2):E46–E52. doi: 10.1055/s-0035-1564268

Table 2 Flow rate (mean, sd) measured using ultrasound and using a transit-time flowmeter. The end column shows the difference in individual animals (median, range).

Ultrasound flow (ml.min−1) Transit-time flow (ml.min−1) p-value Difference %
Separate (II, III) 4.1 (5.0) 5.6 (3.2) 0.16 137 (11–2 900)
Simultaneous (I) 11.9 (10.0) 7.5 (4.6) 0.24 42 (12–1 450)
Rat femoral 1.00 (0.81) 2.2 (0.6) 0.04* 138 (21–2112)
Rat carotid 11.8 (7.4) 9.0 (2.3) 0.23 28 (11–178)
Mouse carotid 1.2 (0.6) 5.2 (3.0) 0.04* 171 (136–2900)

*Statistically significant