Table 6. Autofluorescence and specific fluorescence at peak of emission for Cy3 & Cy5 excited by white light LED (SOLA), PE2 (535nm & 615nm) or PE4000 (555 & 635nm).
Cy3 at 570nm | PE2 535nm | PE4000 555nm | Sola | |||
BK | specific | BK | specific | BK | specific | |
average | 1.989757 | 9.160652 | 1.588668 | 6.795557 | 1.327651 | 5.540776 |
sem | 0.253373 | 0.182925 | 0.014931 | 0.123194 | 0.224278 | 0.105282 |
ratio specific/bk | 4.603906 | 4.277519 | 4.173367 | |||
Cy5 at 670nm | PE2 615nm | PE4000 635nm | Sola | |||
BK | specific | BK | specific | BK | specific | |
average | 0.272805 | 0.919322 | 0.26374 | 1.122252 | 0.128934 | 0.52169 |
sem | 0.01203 | 0.022824 | 0.025742 | 0.02747 | 0.010783 | 0.014581 |
ratio specific/BK | 3.369884 | 4.255147 | 4.046169 |
The specific staining was quantified in thresholded positive cells and 3 regions of interest were drawn manually in unstained tissue for quantification of the background fluorescence (BK).
The average fluorescence intensity (counts/s) was quantified at the peak of fluorescence on images taken at the appropriate wave length by the Nuance camera with Tunable filter (ie 570nm for Cy3 & 670nm for Cy5).
Peaks of excitation for Cy3 & Cy5 are 554nm & 649nm respectively.