Skip to main content
. Author manuscript; available in PMC: 2016 Sep 20.
Published in final edited form as: IEEE Trans Nucl Sci. 2008 Dec 4;55(5):2508–2517. doi: 10.1109/TNS.2008.2004278

TABLE I.

Measured Intrinsic Spatial Resolution of the I-EMCCD Camera With an Optical Pixel1 Size of 96 mm and an 250 mm Thick CsI(Tl) Scintillator

EM Gain WLS Maximum Likelihood Centroiding
FWHM (µm) FWTM (µm) FWHM (µm) FWTM (µm) FWHM (µm) FWTM (µm)
×1000 55.1 115.4 51.9 110.7 51.9 111.9
×200 58.8 116.0 60.1 119.5 56.3 110.1
×20 73.0 124.2 88.7 170.4 71.2 139.5
×5 95.9 165.7 90.5 212.4 108.3 162.2
1

An optical pixel on the DM photocathode is the maximum area that can be projected onto a single pixel on the EMCCD sensor (16µm×16µm)