Published in final edited form as: J Magn Reson Imaging. 2015 Oct 15;43(5):1064–1072. doi: 10.1002/jmri.25077

Table 3.

Inter-platform reproducibility of liver stiffness (LS) and spleen stiffness (SS) measurements between the two platforms and for each observer (Obs. 1 and Obs. 2).

p* ICC CV (%) CR (%) Mean bias (%) BALA
LS Obs. 1 0.363 0.976 9.2 25.4 −3.3 [−28.7, 22.0]
Obs. 2 0.326 0.979 11.5 29.7 −4.2 [−33.9, 25.4]
SS Obs. 1 0.079 0.884 14.4 36.2 −9.2 [−45.4, 26.9]
Obs. 2 0.061 0.912 13.1 35.6 −10.4 [−46.0, 25.1]
*

Paired t-test to compare scanners in terms of mean stiffness.

ICC: intra-class correlation

CV: coefficient of variation (%)

CR: coefficient of reproducibility (%)

BALA: Bland-Altman limits of agreement (%)