Table 4.

Inter- and intra-observer reproducibility of liver stiffness (LS) and spleen stiffness (SS) measurements for each platform.

System # ICC CV (%) CR (%) Mean Bias (%) BALA
Inter-observer
LS 1 0.992 5.4 12.4 0.4 [−12.0, 12.8]
2 0.979 11.2 15.9 1.1 [−14.8, 16.9]
SS 1 0.980 10.0 14.8 5.6 [−9.1, 20.4]
2 0.989 15.7 10.2 4.2 [−5.9, 14.4]
Intra-observer
LS 1 0.999 1.2 2.9 −0.1 [−3.0, 2.7]
2 0.999 1.5 4.2 −0.8 [−5.0, 3.3]
SS 1 0.998 2.1 6.6 −2.0 [−8.6, 4.5]
2 0.998 1.9 3.5 −1.2 [−4.7, 2.2]

System #1: 3.0T GE Discovery MR750, system #2: 1.5T Siemens MAGNETOM Aera

CV: coefficient of variation (%)

CR: coefficient of reproducibility (%)

BALA: Bland-Altman limits of agreement (%)