Table 1.

Imaging parameters for the slow protocol involving 2D UTE single-echo acquisition, 2D IR-UTE single-echo acquisition, 3D UTE single-echo acquisition and 3D IR-UTE single-echo acquisition, and the fast protocol involving 3D dual-echo UTE acquisition and 3D IR-UTE single-echo acquisition with reduced number of TEs (5 radial spokes were acquired per IR preparation in each 3D IR-UTE acquisition).

Sequences TR
(ms)
TE
(ms)
TI
(ms)
FOV
(cm)
Acquisition
Matrix
#spokes #slice Thickness
(mm)
BW
(kHz)
Flip
Angle
Scan time
(min)
Slow

Protocol
2D UTE
Single-Echo
100 0.008; 0.1; 0.2; 0.3; 0.4; 0.6;
0.8; 1; 1.2; 1.6; 2; 3; 4; 6; 8; 10
- 15 192 × 192 191 1 - 250 10° 0.32
2D IR-UTE
Single-Echo
100 0.008; 0.1; 0.2; 0.3; 0.4; 0.6;
0.8; 1; 1.2; 1.6; 2; 3; 4; 6; 8; 10
45 15 128 × 128 191 1 - 250 10° 0.32
3D UTE
Single-Echo
14.5 0.008; 0.1; 0.2; 0.3; 0.4; 0.6;
0.8; 1; 1.2; 1.6; 2; 3; 4; 6; 8; 10
- 15 192 × 192 7449 10 7 250 10° 1.8
3D IR-UTE
Single-Echo
100 0.03; 0.1; 0.2; 0.3; 0.4; 0.6; 0.8;
1; 1.2; 1.6; 2; 3; 4; 6; 8; 10
45 15 128 × 128 6131 10 7 250 10° 2.0
Fast
Protocol
3D UTE
Dual-Echo
14.5 0.03/2.2; 0.2/4.4; 0.4/6.6;
0.6/8.8; 0.8/11
- 15 192 × 192 7449 10 10 250 10° 1.8
3D IR-UTE
Single-Echo
100 0.03; 0.2; 0.4; 0.6; 0.8;
2.2; 4.4; 6.6; 8.8; 11.0
45 15 128 × 128 6131 10 10 250 10° 2.0